Fluke 124 FE 123 & 124 Users Manual - Page 78
Environmental, Safety, Max. Input Voltage Input A and B
View all Fluke 124 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 78 highlights
Fluke 123/124 Users Manual Environmental Environmental MIL-PRF-28800F, Class 2 Temperature Operating 0 to 50 °C (32 to 122 °F) Storage 20 to 60 °C (-4 to 140 °F) Humidity Operating: @0 to 10 °C (32 to 50 °F noncondensing @10 to 30 °C (50 to 86 °F 95% @30 to 40 °C (86 to 104 °F 75% @40 to 50 °C (104 to 122 °F 45% Storage: @-20 to 60 °C (-4 to 140 °F noncondensing Altitude Operating 4.5 km (15 000 feet) Max. Input and Floating Voltage 600 Vrms upto 2 km, linearly derating to 400 Vrms @ 4.5 km Storage 12 km (40 000 feet) Vibration (sinusoidal MIL28800F, class 2, 3.8.4.2, 4.5.5.3.1: Max. 3g Shock MIL28800F, class 2, 3.8.5.1, 4.5.5.4.1: Max. 30g 70 Electromagnetic Compatibility (EMC) Emission EN 50081-1 (1992): EN55022 and EN60555-2 Immunity EN 50082-2 (1992): IEC1000-4-2, -3, -4, -5 (See also Tables 1 to 3) Enclosure Protection IP51, ref: IEC529 Safety Designed for measurements on 600 Vrms Category III Installations, Pollution Degree 2, per: • ANSI/ISA S82.01-1994 • EN61010-1 (1993) (IEC1010-1) • CAN/CSA-C22.2 No.1010.1-92 (including approval) • UL3111-1 (including approval) Max. Input Voltage Input A and B Direct on input or with leads 600 Vrms for derating, see Figure 4-1. With Banana-to BNC Adapter BB120 300 Vrms for derating, see Figure 4-1. Max. Floating Voltage from any terminal to ground 600 Vrms up to 400 Hz