Frequency range 27 MHz to 1 GHz, Trace disturbance with STL120, Table 1, No visible disturbance
View all Fluke 124 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 80 highlights
Fluke 123/124 Users Manual The Fluke 123/124, including standard accessories, conforms with the EEC directive 89/336 for EMC immunity, as defined by IEC1000-4-3, with the addition of the following tables. Trace disturbance with STL120 Table 1 No visible disturbance E= 3 V/m E= 10 V/m Frequency range 10 kHz to 27 MHz Frequency range 27 MHz to 1 GHz 100 mV/div to 500 V/div 100 mV/div to 500 V/div 500 mV/div to 500 V/div 100 mV/div to 500 V/div Table 2 Disturbance less than 10% of full scale E= 3 V/m E= 10 V/m Frequency range 10 kHz to 27 MHz Frequency range 27 MHz to 1 GHz 20 mV/div to 50 mV/div 10 mV/div to 20 mV/div 100 mV/div to 200 mV/div - (-): no visible disturbance Test tool ranges not specified in tables 1 and 2 may have a disturbance of more than 10% of full scale. 72