Fluke 125B/S Calibration Guide - Page 16

Cursor Readout (124B, 125B), Advanced Meter Functions

Page 16 highlights

123B/124B/125B Calibration Manual Continuity (CONT) Beep 30 Ω ±5 Ω) in 50 Ω range Measurement Current 0.5 mA Detection of shorts 1 ms Diode Measurement Voltage @ 0.5 mA 2.8 V @ open circuit 4 V Accuracy 2 % + 5 counts) Measurement Current 0.5 mA Polarity on input A, - on COM Capacitance (CAP) Ranges 50 nF, 500 nF, 5 μF, 50 μF, 500 μF Accuracy 2 % + 10 counts) Full Scale Reading 5000 counts Measurement Current 500 nA to 0.5 mA, increases with increasing ranges Advanced Meter Functions Zero Set Set actual value to reference Fast/Normal/Smooth Meter settling time Fast: 1 s @ 1 μs to 10 ms/div. Meter settling time Normal: 2 s @ 1 μs to 10 ms/div. Meter settling time Smooth: 10 s @ 1 μs to 10 ms/div. AutoHold (on A) Captures and freezes a stable measurement result. Beeps when stable. AutoHold works on the main meter reading, with thresholds of 1 Vpp for AC signals and 100 mV for DC signals. Fixed Decimal Point with attenuation keys Cursor Readout (124B, 125B) Sources A, B Single Vertical Line Average, Min and Max Readout Average, Min, Max and Time from Start of Readout (in ROLL mode, instrument in HOLD) Min, Max and Time from Start of Readout (in RECORDER mode, instrument in HOLD) Harmonics values in POWER QUALITY mode. Dual Vertical Lines Peak-Peak, Time Distance and Reciprocal Time Distance Readout Average, Min, Max and Time Distance Readout (in ROLL mode, instrument in HOLD) Dual Horizontal Lines High, Low and Peak-Peak Readout Rise or Fall Time Transition Time, 0 %-Level and 100 %-Level Readout (Manual or Auto Leveling; Auto Leveling only possible in Single Channel Mode) Accuracy As Oscilloscope Accuracy 8

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54

123B/124B/125B
Calibration Manual
8
Continuity (CONT)
Beep
....................................................................
<(30
Ω
±
5
Ω
) in 50
Ω
range
Measurement Current
.........................................
0.5 mA
Detection of shorts
..............................................
1 ms
Diode
Measurement Voltage
@ 0.5 mA
........................................................
>2.8 V
@ open circuit
.................................................
<4 V
Accuracy
.....................................................................
±
(2 % + 5 counts)
Measurement Current
.........................................
0.5 mA
Polarity
................................................................
+ on input A, - on COM
Capacitance (CAP)
Ranges
................................................................
50 nF, 500 nF, 5
μ
F, 50
μ
F, 500
μ
F
Accuracy
.............................................................
±(2 % + 10 counts)
Full Scale Reading
..............................................
5000 counts
Measurement Current
.........................................
500 nA to 0.5 mA, increases with increasing ranges
Advanced Meter Functions
Zero Set
Set actual value to reference
Fast/Normal/Smooth
Meter settling time Fast: 1 s @ 1
μ
s to 10 ms/div.
Meter settling time Normal: 2 s @ 1
μ
s to 10 ms/div.
Meter settling time Smooth: 10 s @ 1
μ
s to 10 ms/div.
AutoHold (on A)
Captures and freezes a stable measurement result. Beeps when stable. AutoHold works on the main meter reading, with thresholds of
1 Vpp for AC signals and 100 mV for DC signals.
Fixed Decimal Point
..............................................
with attenuation keys
Cursor Readout (124B, 125B)
Sources
A, B
Single Vertical Line
Average, Min and Max Readout
Average, Min, Max and Time from Start of Readout (in ROLL mode
,
instrument in HOLD)
Min, Max and Time from Start of Readout (in RECORDER mode
,
instrument in HOLD)
Harmonics values in POWER QUALITY mode.
Dual Vertical Lines
Peak-Peak, Time Distance and Reciprocal Time Distance Readout
Average, Min, Max and Time Distance Readout (in ROLL mode, instrument in HOLD)
Dual Horizontal Lines
High, Low and Peak-Peak Readout
Rise or Fall Time
Transition Time, 0 %-Level and 100 %-Level Readout (Manual or Auto Leveling; Auto Leveling only possible in Single Channel Mode)
Accuracy
As Oscilloscope Accuracy