Fluke 125B/S Calibration Guide - Page 31

Smooth, Table 4. Volts DC Measurement Verification Points, Sensitivity, Oscilloscope, Range, Meter

Page 31 highlights

Industrial ScopeMeter Performance Verification d. Push . e. Push  to open the INPUT SETTINGS A menu. f. Push  to open the INPUT SETTINGS B menu. g. Make these selections: • Select Coupling: DC • Readings: Smooth • Noise filter: ON 7. For each sensitivity value listed in Table 4: a. On the Test Tool, push  or  to set the Input A and Input B sensitivity. The range is listed in the second column. b. Set up the Calibrator: i. Source the DC voltage listed in Table 4. ii. Push . c. Compare the main reading to the limits in Table 4. 8. Set the Calibrator to 0 V (zero). Table 4. Volts DC Measurement Verification Points Sensitivity (Oscilloscope) Range[1] (Meter) Calibrator Output V DC 5 mV/div 500 mV 15 mV 10 mV/div 500 mV 30 mV 20 mV/div 500 mV 60 mV 50 mV/div 500 mV 150 mV 100 mV/div 500 mV 300 mV 500 mV 200 mV/div 500 mV -500 mV 0 mV 500 mV/div 5 V 1.5 V 1 V/div 5 V 3 V 5 V 2 V/div 5 V -5 V 0 V 5 V/div 50 V 15 V 10 V/div 50 V 30 V 50 V 20 V/div 50 V -50 V 0 V 50 V/div 500 V 150 V 100 V/div 500 V 300 V [1] The 500 V and 750 V range is tested in Input A and B High Voltage AC/DC Accuracy Test on page 28. [2] OL (overload) shows on the display occasionally when there is calibrator noise. Input A-B DC Limit 014.4 to 015.6[2] 029.3 to 030.7[2] 059.2 to 060.8 148.7 to 151.3 298.0 to 302.0 497.0 to 503.0 -497.0 to -503.0 -000.5 to + 000.5 1.487 to 1.513 2.980 to 3.020 4.970 to 5.030 -4.970 to -5.030 -0.005 to +0.005 14.87 to 15.13 29.80 to 30.20 49.70 to 50.30 -49.70 to -50.30 -00.05 to +00.05 148.7 to 151.3 298.0 to 302.0 23

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54

Industrial ScopeMeter
Performance Verification
23
d.
Push
.
e.
Push
to open the INPUT SETTINGS A menu.
f.
Push
to open the INPUT SETTINGS B menu.
g.
Make these selections:
Select Coupling:
DC
• Readings:
Smooth
Noise filter:
ON
7.
For each sensitivity value listed in Table 4:
a.
On the Test Tool, push
or
to set the Input A and Input B sensitivity.
The range is listed in the second column.
b.
Set up the Calibrator:
i.
Source the DC voltage listed in Table 4.
ii. Push
.
c.
Compare the main reading to the limits in Table 4.
8.
Set the Calibrator to 0 V (zero).
Table 4. Volts DC Measurement Verification Points
Sensitivity
(Oscilloscope)
Range
[1]
(Meter)
Calibrator Output
V DC
Input A-B DC Limit
5 mV/div
500 mV
15 mV
014.4 to 015.6
[2]
10 mV/div
500 mV
30 mV
029.3 to 030.7
[2]
20 mV/div
500 mV
60 mV
059.2 to 060.8
50 mV/div
500 mV
150 mV
148.7 to 151.3
100 mV/div
500 mV
300 mV
298.0 to 302.0
200 mV/div
500 mV
500 mV
497.0 to 503.0
-500 mV
-497.0 to -503.0
0 mV
-000.5 to + 000.5
500 mV/div
5 V
1.5 V
1.487 to 1.513
1 V/div
5 V
3 V
2.980 to 3.020
2 V/div
5 V
5 V
4.970 to 5.030
-5 V
-4.970 to -5.030
0 V
-0.005 to +0.005
5 V/div
50 V
15 V
14.87 to 15.13
10 V/div
50 V
30 V
29.80 to 30.20
20 V/div
50 V
50 V
49.70 to 50.30
-50 V
-49.70 to -50.30
0 V
-00.05 to +00.05
50 V/div
500 V
150 V
148.7 to 151.3
100 V/div
500 V
300 V
298.0 to 302.0
[1]
The 500 V and 750 V range is tested in
Input A and B High Voltage AC/DC Accuracy Test
on page 28.
[2]
OL (overload) shows on the display occasionally when there is calibrator noise.