Fluke 125B/S Calibration Guide - Page 33
Input A and Input B AC Input Coupling Test, Input A and Input B Volts Peak Measurements Test
View all Fluke 125B/S manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 33 highlights
Industrial ScopeMeter Performance Verification Input A and Input B AC Input Coupling Test To test the Input A and B ac input coupling: 1. Connect the Test Tool to the Calibrator. See Figure 6. 2. Set up the Test Tool: a. Push . b. Push to open the INPUT SETTINGS A menu and make these selections: • Select Coupling: AC • Readings: Smooth • Noise filter: OFF c. Push . d. Push to open the INPUT SETTINGS B menu and make these selections: • Coupling: AC • Readings: Smooth • Noise filter: OFF 3. For each sensitivity value listed in Table 6: a. On the Test Tool, push or to set the Input A and Input B sensitivity. b. Set up the Calibrator: i. Supply the voltage and frequency listed in Table 6. ii. Push . c. Compare the Test Tool reading to the limit in Table 6. d. Set the Calibrator to 0 V (zero). Table 6. Input A and Input B AC Input Coupling Verification Points Calibrator Output V rms 500 mV 500 mV 500 mV Calibrator Frequency 10 Hz 33 Hz 60 Hz >344.0 >469.0 >486.5 Limit A-B Input A and Input B Volts Peak Measurements Test XW Warning To prevent possible electrical shock, fire, or personal injury, make sure that the calibrator is in standby mode before making any connection between the calibrator and the Test Tool. Dangerous voltages will be present on the calibration source and connecting cables during the following steps. To test the Input A and Input B volts peak measurements: 1. Connect the Test Tool to the Calibrator. See Figure 6. 2. Setup Peak for Input A on the Test Tool: a. Push to select auto ranging (AUTO shows at the top of the display). b. Push . c. Push to open the INPUT A MEASUREMENTS menu. d. Select Peak and push to open the Peak submenu. e. Select Peak Max-Min. 25