Fluke 125B/S Calibration Guide - Page 39
Diode Function Test, Capacitance Measurements Test, DIODE
View all Fluke 125B/S manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 39 highlights
Industrial ScopeMeter Performance Verification 3. Set up the Calibrator: a. Supply 25 Ω b. Use the "COMP 2 wire" mode. c. Push . 4. Make sure the beeper sounds continuously. 5. On the Calibrator, supply 35 Ω. 6. Make sure the beeper does not sound. Diode Function Test To test the diode function: 1. Connect the Test Tool to the Calibrator. See Figure 9. 2. Setup the Test Tool: a. Push . b. Push to open the INPUT A MEASUREMENTS menu. c. Select DIODE. 3. Set up the Calibrator: a. Supply 1 kΩ. b. Use the Calibrator "COMP 2 wire" mode. c. Push . 4. Verify the main reading is between 0.425 V and 0.575 V. 5. On the Calibrator, supply 1 V DC. 6. Verify the main reading is between 0.975 V and 1.025 V. Capacitance Measurements Test To test the capacitance measurement function: 1. Connect the Test Tool to the Calibrator. See Figure 9. 2. Set up the Test Tool: a. Push . b. Push to open the INPUT A MEASUREMENTS menu. c. Select CAP. 3. Push and to select manual ranging. 4. For each Range and Output value in Table 11: a. Set up the Calibrator: i. Supply the Range and Output. ii. Use the Calibrator "COMP OFF" mode. iii. Push . b. Compare the Input A main reading to the limit listed in Table 11. Table 11. Input A and Input B AC Input Coupling Verification Points 0.56 nF 0.5 nF 5 nF 500 nF 0.05 μF 5 μF 0.5 μF 50 μF 5 μF 500 μF Calibrator Output 50 nF 50 nF 500 nF 500 nF 5 μF 5 μF 50 μF 50 μF 500 μF 500 μF Range 0.45 to 0.67 48.90 to 51.10 3.9 to 6.1 489.0 to 511.0 0.039 to 0.061 4.890 to 5.110 0.39 to 0.61 48.90 to 51.10 3.9 to 6.1 489.0 to 511.0 Limit 31