Fluke 125B/S Calibration Guide - Page 44
Position, Input A and B, Pulse Adjust Input B
View all Fluke 125B/S manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 44 highlights
123B/124B/125B Calibration Manual 7. Connect the calibrator Scope Output to both CH A and CH B without the 50 Ω terminator. 8. Set up the Calibrator: a. Supply a 1 kHz square wave (Output SCOPE, Mode VOLT). b. Set the Calibration step in Table 14. c. Push . 9. On the Test Tool: a. Push to select the first Calibration Step in Table 14. b. Push to start the calibration. c. Wait until READY shows as the calibration status on the display. d. Push to select the next calibration step. 10. Repeat steps 8 through 9 for each Calibration Step in Table 14. 11. Put the Calibrator in Standby. 12. Continue to the next section. Table 14. HF Gain Calibration Points Slow Calibration Step LF-Gain- (CL 0330) 25 V LF-Gain (CL 0335) 50 V [1] After starting the first step in this table cell, other steps are done automatically. Calibrator Setting [1] (1 kHz, Scope Mode EDGE) Position, Input A and B To do the position calibration: 1. Push to select calibration step, Position (CL 0400):IDLE. 2. Disconnect all connections to the Test Tool inputs. 3. Push to start the calibration. The Position (CL0400) to (CL0414) calibrates. 4. Wait until Position (CL 0414):READY shows on the display. 5. Continue to the next section. Pulse Adjust Input B To do the Pulse Adjust Input A calibration: 1. Push to select calibration step Volt Zero (CL 0420):IDLE. 2. Connect 50 Ω Terminators to the A and B inputs (BB120 and TRM50). 3. Push to start the calibration. 4. Wait until Volt Zero (CL 0434):READY shows on the display. 5. Continue to the next section. 36