Fluke 1503 FE 1503-1507 Users Manual - Page 29

AC/DC Voltage Measurement, Range, Resolution, Hz to 400 Hz, of Rdg + Digits - specifications

Page 29 highlights

Insulation Testers Specifications IP Rating IP40 Altitude Operating: 2000 m CAT IV 600 V, 3000 m CAT III 600 V Non Operating (Storage): 12,000 m Over-Range Capability 110% of range Compliance to EN 61557 IEC61557-1, IEC61557-2, IEC61557-4, IEC61557-10 Model 1503 Accessories TL224 Leads TP74 Probes Clips PN 1958654 (red) and PN 1958646 (black) Holster Model 1507 Accessories TL224 Leads TP74 Probes Clips PN 1958654 (red) and PN 1958646 (black) Holster Remote Probe AC/DC Voltage Measurement Accuracy Range Resolution 50 Hz to 400 Hz ± (% of Rdg + Digits) 600.0 V 0.1 V + (2 % + 3) Input Impedance 3 MΩ (nominal),

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38

Insulation Testers
Specifications
21
IP Rating
..............................................................
IP40
Altitude
................................................................
Operating: 2000 m CAT IV 600 V, 3000 m CAT III 600 V
Non Operating (Storage): 12,000 m
Over-Range Capability
........................................
110% of range
Compliance to EN 61557
....................................
IEC61557-1, IEC61557-2, IEC61557-4, IEC61557-10
Model 1503 Accessories
.....................................
TL224 Leads
TP74 Probes
Clips PN 1958654 (red) and PN 1958646 (black)
Holster
Model 1507 Accessories
.....................................
TL224 Leads
TP74 Probes
Clips PN 1958654 (red) and PN 1958646 (black)
Holster
Remote Probe
AC/DC Voltage Measurement
Accuracy
Range
Resolution
50 Hz to 400 Hz
±
(% of Rdg + Digits)
600.0 V
0.1 V
+
(2 % + 3)
Input Impedance
..................................................
3 M
(nominal), <100 pF
Common Mode Rejection Ratio
(1 k
unbalanced)
...............................................
>60 dB at dc, 50 or 60 Hz
Overload Protection
.............................................
600 V rms or dc