Fluke 1503 FE 1503-1507 Users Manual - Page 31
Model 1507
View all Fluke 1503 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 31 highlights
Model 1507 Output Voltage 50 V (0 % to + 20 %) 100 V (0 % to + 20 %) 250 V (0 % to + 20 %) 500 V (0 % to + 20 %) 1000 V (0 % to + 20 %) Display Range 0.01 to 20.00 MΩ 20.0 to 50.0 MΩ 0.01 to 20.00 MΩ 20.0 to 100.0 MΩ 0.01 to 20.00 MΩ 20.0 to 200.0 MΩ 0.01 to 20.00 MΩ 20.0 to 200.0 MΩ 200 to 500 MΩ 0.1 to 200.0 MΩ 200 to 2000 MΩ 2.0 to 10.0 GΩ Resolution 0.01 MΩ 0.1 MΩ 0.01 MΩ 0.1 MΩ 0.01 MΩ 0.1 MΩ 0.01 MΩ 0.1 MΩ 1 MΩ 0.1 MΩ 1 MΩ 0.1 GΩ Insulation Testers Specifications Test Current 1 mA @ 50 kΩ Accuracy ± (% of Rdg + Digits) ± (3 % + 5) 1 mA @ 100 kΩ ± (3 % + 5) 1 mA @ 250 kΩ ± (1.5 % + 5) 1 mA @ 500 kΩ ± (1.5 % + 5) 1 mA @ 1 MΩ ± (1.5 % + 5) ± (10 % + 3) 23
Insulation Testers
Specifications
23
Model 1507
Output Voltage
Display Range
Resolution
Test Current
Accuracy
±
(% of Rdg + Digits)
0.01 to 20.00 M
Ω
0.01 M
Ω
50 V
(0 % to + 20 %)
20.0 to 50.0 M
Ω
0.1 M
Ω
1 mA @ 50 k
Ω
±
(3 % + 5)
0.01 to 20.00 M
Ω
0.01 M
Ω
100 V
(0 % to + 20 %)
20.0 to 100.0 M
Ω
0.1 M
Ω
1 mA @ 100 k
Ω
±
(3 % + 5)
0.01 to 20.00 M
Ω
0.01 M
Ω
250 V
(0 % to + 20 %)
20.0 to 200.0 M
Ω
0.1 M
Ω
1 mA @ 250 k
Ω
±
(1.5 % + 5)
0.01 to 20.00 M
Ω
0.01 M
Ω
20.0 to 200.0 M
Ω
0.1 M
Ω
500 V
(0 % to + 20 %)
200 to 500 M
Ω
1 M
Ω
1 mA @ 500 k
Ω
±
(1.5 % + 5)
0.1 to 200.0 M
Ω
0.1 M
Ω
200 to 2000 M
Ω
1 M
Ω
±
(1.5 % + 5)
1000 V
(0 % to + 20 %)
2.0 to 10.0 G
Ω
0.1 G
Ω
1 mA @ 1 M
Ω
±
(10 % + 3)