Fluke 225C Getting Started Guide - Page 12
Recorder Functions, Fieldbus Measurements (Bushealth), Selecting the Bus Type
View all Fluke 225C manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 12 highlights
Fluke 19xC-2x5C Getting Started Recorder Functions First, choose a measurement in Scope or Meter mode. RECORDER ANALYZE Show the recorder functions. Select a recorder function (Scope Trendplot, Scope Record, or Meter Trendplot), then press to display the recorder key labels: Start (RUN) or stop (STOP) recording. Show the recorder options. Reference: time reference is time of day (Time of Day) or time from start (From Start). In Scope Record: Display Glitches: glitch detection on, or 10 kHz filter on. Mode: Single Sweep : stores samples until memory is full, then stops. Mode: Continuous : stores samples continuously; deletes first samples if memory is full. Mode: on Ext. ... : start (Start), stop (Stop) or continue (Run) recording if triggered via the meter input (Ext.). Run requires one trigger per division in VIEW ALL mode. 8 In Trend Plot: Reading 1 (2)... (Scope) or Measurement... (Meter) show the measurement function menu. VIEW ALL : see all stored samples. NORMAL : see the most recent 9 divisions. EXIT : exit the recorder mode. Fieldbus Measurements (Bushealth) The Bushealth function is standard available in the Fluke 215C-225C. Selecting the Bus Type ANALYZE Show the function selection screen. Select Bushealth , then press . Highlight the required bus type, the press . Dots ... behind a bus type indicate that a submenu will follow on this selection. The test tool shows the bushealth screen. Show information on how to connect the test tool to the bus (WIRING INFO). Show the bus signal eye pattern (waveform).