Fluke 225C Getting Started Guide - Page 13
Bushealth Screen, Setting up the Limits, Result not OK.
View all Fluke 225C manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 13 highlights
Toggle between scope screen (OFF) and bushealth screen (ON). Bushealth Screen The following icons are used to indicate the bus measurement status: bus activity indicators: Q (filled) : voltage measured { (open) : no voltage measured ÑÑ(blinking) : activity { { (both open) : no activity Busy, the tester is measuring/processing data. No reading available. Result OK . Warning. Result on edge of limits Result not OK. Getting Started Fieldbus Measurements (Bushealth) Setting up the Limits To set up the test limits of the current bus type, do the following: From the bushealth screen select SETUP LIMITS... For RS-485 only: highlight Manchester or NRZ (default is Manchester). Press to select the encoding type Select the signal property for which you want to change the limits. Select the limit you want to change, LOW (low limit), HIGH (high limit), WARNING (warn limit) Change the limit. Continue with all limits to be changed. Save the limits and return to the bushealth screen. Note: Changed limits are marked by a * (asterisk). 9