Fluke 83V FE 83V & 87V Users Manual - Page 29
Measuring Capacitance, Measuring Capacitance
View all Fluke 83V manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 29 highlights
Measuring Capacitance WCaution To avoid possible damage to the Meter or to the equipment under test, disconnect circuit power and discharge all high-voltage capacitors before measuring capacitance. Use the dc voltage function to confirm that the capacitor is discharged. The Meter's capacitance ranges are 10.00 nF, 100.0 nF, 1.000 µF, 10.00 µF, 100.0 µF, and 9999 µF. To measure capacitance, set up the Meter as shown in Figure 6. To improve the accuracy of measurements less than 1000 nF, use the relative (REL) mode to subtract the residual capacitance of the Meter and leads. Note If too much electrical charge is present on the capacitor being tested, the display shows "diSC". Making Measurements Select Capacitance 87 V TRUE RMS MULTIMETER MIN MAX Peak MIN MAX 4½ DIGITS 1 Second ˚C/˚F RANGE REL mV V AutoHOLD Hz % mA A LO V A OFF A mA A COM V 10A MAX FUSED 400mA FUSED + Figure 6. Measuring Capacitance aom10f.eps 21