Fluke 83V FE 83V & 87V Users Manual - Page 58
Table 17. Frequency Counter Sensitivity and Trigger Levels, Minimum Sensitivity RMS Sine wave
View all Fluke 83V manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 58 highlights
80 Series V Users Manual Table 17. Frequency Counter Sensitivity and Trigger Levels Input Range1 Minimum Sensitivity (RMS Sine wave) 5 Hz - 20 kHz 0.5 Hz - 200 kHz 600 mV dc 600 mV ac 6 V 60 V 600 V 1000 V 70 mV (to 400 Hz) 150 mV 0.3 V 3 V 30 V 100 V 70 mV (to 400 Hz) 150 mV 0.7 V 7 V (≤140 kHz) 70 V (≤14.0 kHz) 200 V (≤1.4 kHz) Duty Cycle Range 0.0 to 99.9 % Within ± (0.2% per kHz + 0.1 %) for rise times < 1 µs. 1. Maximum input for specified accuracy = 10X Range or 1000 V. Approximate Trigger Level (DC Voltage Function) 40 mV 1.7 V 4 V 40 V 100 V Accuracy 50
80 Series V
Users Manual
50
Table 17. Frequency Counter Sensitivity and Trigger Levels
Minimum Sensitivity (RMS Sine wave)
Approximate Trigger Level
Input Range
1
5 Hz - 20 kHz
0.5 Hz - 200 kHz
(DC Voltage Function)
600 mV dc
600 mV ac
6 V
60 V
600 V
1000 V
70 mV (to 400 Hz)
150 mV
0.3 V
3 V
30 V
100 V
70 mV (to 400 Hz)
150 mV
0.7 V
7 V (
≤
140 kHz)
70 V (
≤
14.0 kHz)
200 V (
≤
1.4 kHz)
40 mV
1.7 V
4 V
40 V
100 V
Duty Cycle Range
Accuracy
0.0 to 99.9 %
Within
±
(0.2% per kHz + 0.1 %) for rise times < 1
µ
s.
1.
Maximum input for specified accuracy = 10X Range or 1000 V.