P3000 Series Users Manual Connections Fit the device under test (DUT) to the test port using the method described below: W Warning DO NOT use Teflon/PTFE tape on these connections, as this will prevent correct sealing. The Gauge Adapter sealing system is designed for hand-tight sealing up to 20,000 psi / 1,400 bar-wrenches or similar tools are not required - over tightening can cause damage to threads or sealing faces. Before connection, ensure that there is an O-ring fitted to the test port. Check that the sealing face of the device to be fitted is clean and undamaged, as scratches or dents can form leak-paths. Note The thread on the test port, and the lower part of the gauge adapters is LEFT-HANDED. The following procedure details the correct method for mounting devices using these adapters: 1. Screw the appropriate gauge adapter fully on to the instrument to be tested. Figure 2-1. Screw on Gauge Adapter 2. Screw assembly down COUNTER-CLOCKWISE on to test port. Note Hand-tight is sufficient; ensure that the bottom face contacts the O-ring on the test port. gjn006.bmp 2-2 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41

P3000 Series
Users Manual
2-2
Connections
Fit the device under test (DUT) to the test port using the method described below:
W
Warning
DO NOT use Teflon/PTFE tape on these connections, as this
will prevent correct sealing. The Gauge Adapter sealing system
is designed for hand-tight sealing up to 20,000 psi / 1,400
bar-wrenches or similar tools are not required — over
tightening can cause damage to threads or sealing faces.
Before connection, ensure that there is an O-ring fitted to the
test port.
Check that the sealing face of the device to be fitted is clean
and undamaged, as scratches or dents can form leak-paths.
Note
The thread on the test port, and the lower part of the gauge adapters is
LEFT-HANDED. The following procedure details the correct method for
mounting devices using these adapters:
1.
Screw the appropriate gauge adapter fully on to the instrument to be tested.
gjn006.bmp
Figure 2-1. Screw on Gauge Adapter
2.
Screw assembly down COUNTER-CLOCKWISE on to test port.
Note
Hand-tight is sufficient; ensure that the bottom face contacts the O-ring on
the test port.