Hoveround MPV5 User Manual - Page 57

VSI Controller, bars Rapid Flash, bars, Three Second Blink, Speed Bars Cycling, Speed Bars Flashing - motor

Page 57 highlights

Hoveround-MPV5ManualRevM 1/17/14 10:37 AM Page 57 If a fault occurs, you can find out what has happened by counting the number of bars on the battery gauge that are flashing. VSI Controller 1 bar: Low battery or a bad connection. Try recharging the battery. 2 bars: Bad connection to left motor. 3 bars: Short circuit, left motor to battery. 4 bars: Bad connection to right motor. 5 bars: Short circuit, right motor to battery. 6 bars: Check if battery charger is still connected to controller or controller fault. 7 bars: Joystick fault. 8 bars: Controller fault. 9 bars: Problem with brake circuit. Check that manual brake release levers are in the drive position and power seat lift is lowered (if equipped). 10 bars: An excessive voltage to the control system. This is usually caused by a poor battery connection. 10 bars Rapid Flash: Powered up when joystick deflected. Restart with joystick in neutral. 10 bars, Three Second Blink: Control is in sleep mode. Press OFF/ON twice to restart. Speed Bars Cycling: Controller locked. Speed Bars Flashing: Power seat lift is inhibiting drive. Lower the seat. D82008753M 10/13 Chapter 9 57

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68

D82008753M
10/13
Chapter 9
57
If a fault occurs, you can find out what has happened by counting
the number of bars on the battery gauge that are flashing.
VSI Controller
1 bar:
Low battery or a bad connection.
Try recharging the battery.
2 bars:
Bad connection to left motor.
3 bars:
Short circuit, left motor to battery.
4 bars:
Bad connection to right motor.
5 bars:
Short circuit, right motor to battery.
6 bars:
Check if battery charger is still connected
to controller or controller fault.
7 bars:
Joystick fault.
8 bars:
Controller fault.
9 bars:
Problem with brake circuit. Check that manual brake
release levers are in the drive position and power seat
lift is lowered (if equipped).
10 bars:
An excessive voltage to the control system. This is
usually caused by a poor battery connection.
10 bars Rapid Flash:
Powered up when joystick deflected.
Restart with joystick in neutral.
10 bars, Three Second Blink:
Control is in sleep mode.
Press OFF/ON twice to restart.
Speed Bars Cycling:
Controller locked.
Speed Bars Flashing:
Power seat lift is inhibiting drive.
Lower the seat.
Hoveround-MPV5ManualRevM
1/17/14
10:37 AM
Page 57