Invacare IH6077A/IH61 Owners Manual - Page 25
Replacing the Release Handle and Cable Assembly, Model IH6077A Only
View all Invacare IH6077A/IH61 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 25 highlights
SECTION 5-MAINTENANCE 8. Turn cylinder 1/2‐turn in the opposite direction of STEP 7. 9. Reassemble by reversing STEPS 1‐6. Mounting Screw Cable Back Frame Gas Spring Recline Cylinder Cable Mounting Bolt DETAIL "A" LOOSENING THE LOCKNUT Locknut Back Frame Retaining Ring Gas Spring Recline Cylinder Clevis Pin Locknut DETAIL "B" TIGHTENING THE CABLE Cylinder Clevis Pin Cable Bottom Bracket FIGURE 5.6 Adjusting Gas Spring Recline Cylinder Assembly - Model IH6077A Only Replacing the Release Handle and Cable Assembly Model IH6077A Only NOTE: For this procedure, refer to FIGURE 5.7 on page 26. 1. With one hand, squeeze release lever up toward mounting bracket and remove anchor end of cable assembly from slot in release lever (Detail ʺBʺ ). CAUTION When removing the cable from the mounting bracket, use care to pull straight out of "V" notch. DO NOT remove by bending the cable. Damage to the cable may occur. 2. Remove cable assembly from ʺVʺ notch of the mounting bracket (Detail ʺBʺ). 3. Remove the mounting screw that secures the cable to the back frame. 4. Remove the mounting screw which secures the release handle to the push handle (Detail ʺAʺ). 5. Secure new release handle to the push handle with the existing mounting screw. Securely tighten (Detail ʺAʺ). Part No. 1141494 25 Invacare Mobile Recliners