SECTION 5-MAINTENANCE Replacing Plunger Release Knob NOTE: For this procedure, refer to FIGURE 5.10 1. Remove glide from end of frame. 2. Remove retaining ring from end of plunger pin. 3. Slide plunger release knob off the plunger pin. 4. Lift plunger pin and spring out of the frame tube. Plunger Pin Frame 5. Insert new plunger pin and spring into the frame so that grooved end of pin extends below the frame tube. 6. Place plunger release knob over the plunger pin. 7. Snap the retaining ring into groove of the plunger pin. 8. Replace the glide into the end of the frame. Spring Glide Plastic Keeper Plunger Release Knob Frame Tube Retaining Ring FIGURE 5.10 Replacing Plunger Release Knob Removing/Installing the Side Panel NOTE: For this procedure, refer to FIGURE 5.11. Removing Side Panels 1. Press the push buttons near the bottom of the right side panel. 2. Hold the push buttons in while pulling the panel outward. 3. Repeat STEPS 1 and 2 for the other side panel. Installing Side Panel 1. Slip the top of the right side panel into the mounting tab. 2. Push the side panel upward until the side panel fit into the side frame. 3. Pop the side panel into position when the push buttons are visible, through the push pin holes, near the bottom of the side frame. Side Panel Push Buttons FIGURE 5.11 Removing/Installing the Side Panel Invacare Mobile Recliners 28 Part No. 1141494