Stitch Adjustment  Remove all the document return guides to prevent the calibration target getting stuck during forward/rewind motion required for calibration.  Clean the scanner.  Insert the calibration target (it must be clean and in good condition) into the scanner face up, black end first.  On the scanner UI, use the left /right buttons as necessary to select the Settings and press OK.  Select Auto-Stitch and press OK. (Auto-Stitch will not start until calibration sheet is loaded and in the correct orientation - Face up, black end first).  The calibration target will be scanned back & forth until complete. Please wait until the scanner returns to the Home screen. This process will take a few minutes. Return the calibration target to its protective cover for safe keeping. Version 1.0 38 Global Scanning UK Ltd © 2016

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50

Version 1.0
38
Global Scanning UK Ltd © 2016
Stitch Adjustment
Remove all the document return guides to prevent the calibration target getting stuck during
forward/rewind motion required for calibration.
Clean the scanner.
Insert the calibration target (it must be clean and in good condition) into the scanner face up,
black end first.
On the scanner UI, use the left /right buttons as necessary to select the Settings and press OK.
Select Auto-Stitch and press OK. (Auto-Stitch will not start until calibration sheet is loaded and in
the correct orientation - Face up, black end first).
The calibration target will be scanned back & forth until complete. Please wait until the scanner
returns to the Home screen. This process will take a few minutes. Return the calibration target
to its protective cover for safe keeping.