1550B Users Manual Specifications General Specifications Display Power Charger Input (AC) Dimensions (H x W x L) Weight Temperature (operating) Temperature (storage) Humidity Altitude Enclosure Sealing Input Overload Protection Electromagnetic Compatibility Certifications Safety Compliance Typical Battery Charge Capability Note At temperature extremes, the battery needs to be charged more frequently. 75 mm x 105 mm 12 V lead-acid rechargeable battery. Yuasa NP2.8-12 85 V to 250 V ac 50/60 Hz 20 VA This Class II (double insulated) instrument is supplied with a Class 1 (grounded) power cord. The protective earth terminal (ground pin) is not connected internally. The extra pin is for added plug retention only. 170 mm x 242 mm x 330 mm (6.7 in. x 9.5 in. x 13.0 in.) 3.6 kg (7.94 lbs.) −20 oC to 50 oC (−4 oF to 122 oF) −20 oC to 65 oC (−4 oF to 149 oF) 80 % to 31 oC decreasing linearly to 50 % at 50 oC 2000 m IP40 600 V AC EN 61326 P s ) EN 61010, EN 61557 Parts 1 and 2 IEC 61010-1, CAT III V 600, Pollution Degree 2 Test Voltages Number of Tests 250 V 4138 500 V 3913 1 kV 3462 2.5 kV 2043 5 kV 1000 22

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36

1550B
Users Manual
22
Specifications
General Specifications
Display
75 mm x 105 mm
Power
12 V lead-acid rechargeable battery.
Yuasa NP2.8-12
Charger Input (AC)
85 V to 250 V ac 50/60 Hz 20 VA
This Class II (double insulated) instrument is
supplied with a Class 1 (grounded) power
cord. The protective earth terminal (ground
pin) is not connected internally. The extra pin
is for added plug retention only
.
Dimensions (H x W x L)
170 mm x 242 mm x 330 mm
(6.7 in. x 9.5 in. x 13.0 in.)
Weight
3.6 kg (7.94 lbs.)
Temperature (operating)
20
o
C to 50
o
C (
4
o
F to 122
o
F)
Temperature (storage)
20
o
C to 65
o
C (
4
o
F to 149
o
F)
Humidity
80 % to 31
o
C decreasing linearly to
50 % at 50
o
C
Altitude
2000 m
Enclosure Sealing
IP40
Input Overload Protection
600 V AC
Electromagnetic Compatibility
EN 61326
Certifications
P
s
)
Safety Compliance
EN 61010, EN 61557 Parts 1 and 2
IEC 61010-1, CAT III V 600,
Pollution Degree 2
Test Voltages
Number of Tests
250 V
4138
500 V
3913
1 kV
3462
2.5 kV
2043
Typical Battery Charge
Capability
Note
At temperature extremes, the
battery needs to be charged
more frequently.
5 kV
1000