MegOhmMeter Specifications Electrical Specifications The meter's accuracy is specified for 1 year after calibration at operating temperatures of 0 °C to 35 °C. For operating temperatures outside the range (−20 °C to 0 °C and 35 °C to 50 °C), add ± .25 % per °C, except on the 20 % bands add ± 1 % per °C. Insulation Test Voltage (DC) Insulation Resistance Range Accuracy (+/- reading) 250 V 50 GΩ 500 V 100 GΩ 1000 V 200 GΩ 2500 V 500 GΩ 5000 V 1 TΩ Bar graph range: Insulation test voltage accuracy: Induced ac mains current rejection: unspecified 5 % 20 % unspecified unspecified 5 % 20 % unspecified unspecified 5 % 20 % unspecified unspecified 5 % 20 % unspecified unspecified 5 % 20 % unspecified 0 to 1 TΩ -0 %, +10 % at 1 mA load current 2 mA maximum Charging rate for capacitive load: 5 seconds per µF Discharge rate for capacitive load 1.5 s/µF Leakage current measurement Range 1 nA to 2 mA Accuracy +/- (5 % + 2 nA) Capacitive measurement Range 0.01 µF to 15.00 µF Accuracy +/- (15 % of reading + 0.03 µF) Timer Range 0 to 99 minutes Live circuit warning Resolution Setting: 1 minute Indication: 1 second Warning range: 30 V to 660 V ac/dc, 50/60 Hz Voltage accuracy: +/- (5 % + 2 V) 23

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36

MegOhmMeter
Specifications
23
Electrical Specifications
The meter’s accuracy is specified for 1 year after calibration at operating
temperatures of 0
°
C to 35
°
C. For operating temperatures outside the range
(
20
°
C to 0
°
C and 35
°
C to 50
°
C), add ± .25 % per °C, except on the 20 %
bands add ± 1 % per °C.
Insulation
Test Voltage (DC)
Insulation Resistance Range
Accuracy (+/- reading)
250 V
<200 k
200 k
to 5 G
5 G
to 50 G
>50 G
unspecified
5 %
20 %
unspecified
500 V
<200 k
200 k
to 10 G
10 G
to 100 G
>100 G
unspecified
5 %
20 %
unspecified
1000 V
<200 k
200 k
to 20 G
20 G
to 200 G
>200 G
unspecified
5 %
20 %
unspecified
2500 V
<200 k
200 k
to 50 G
50 G
to 500 G
>500 G
unspecified
5 %
20 %
unspecified
5000 V
<200 k
200 k
to 100 G
100 G
to 1 T
>1 T
unspecified
5 %
20 %
unspecified
Bar graph range:
Insulation test voltage accuracy:
Induced ac mains current rejection
:
Charging rate for capacitive load:
0 to 1 T
-0 %, +10 % at 1 mA load current
2 mA maximum
5 seconds per
µ
F
Discharge rate for capacitive load
1.5 s/μF
Leakage current measurement
Range
Accuracy
1 nA to 2 mA
+/- (5 % + 2 nA)
Capacitive measurement
Range
Accuracy
0.01
µ
F to 15.00
µ
F
+/- (15 % of reading + 0.03
µ
F)
Timer
Range
Resolution
0 to 99 minutes
Setting: 1 minute
Indication: 1 second
Live circuit warning
Warning range: 30 V to 660 V ac/dc, 50/60 Hz
Voltage accuracy: +/- (5 % + 2 V)