Fluke 1577 FE 1587-1577 Users Manual - Page 16
Buttons, Table 2. Rotary Switch Selections cont., Switch, Position, Measurement Function
View all Fluke 1577 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 16 highlights
1587/1577 Users Manual Table 2. Rotary Switch Selections (cont.) Switch Position X O (1587 and 1587T only) c a INSULATION Measurement Function Continuity test. Beeper turns on at 100 Ω. Diode test. There is no ranging in this function. Displays 0L above 6.600 V. AC mA from 3.00 mA to 400 mA (600 mA overload for 2 minutes maximum). DC mA from 0.01 mA to 400 mA (600 mA overload for 2 minutes maximum). Ohms from 0.01 MΩ to 2 GΩ. Performs insulation test with 50, 100, 250, 500 (default), and 1000 V source on the 1587 or 500 (default) and 1000 V source on the 1577 or 50 V (default) and 100 V on the 1587T. The last selected high voltage setting is retained in memory when the Meter is turned off. Press the blue button to activate smoothing during insulation testing (1587 only). Buttons Use the buttons to activate features that augment the function selected with the rotary switch. The buttons are shown in Figure 2 and described in Table 3. Figure 2. Buttons bav03f.eps 6