DC and AC Current Measurement, Range, Resolution, Accuracy 1587 and 1587T, of Rdg+Digits
View all Fluke 1577 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 43 highlights
Insulation Multimeters Electrical Specifications DC and AC Current Measurement Range AC 45 Hz to 1000 Hz 400 mA 60 mA 400 mA DC 60 mA [1] 1 kHz bandwidth. Resolution 0.1 mA 0.01 mA Accuracy 1587 and 1587T ± (% of Rdg+Digits) + (1.5 % + 2) [1] + (1.5 % + 2) [1] Accuracy 1577 ± (% of Rdg+Digits) + (2 % + 2) [1] + (2 % + 2) [1] 0.1 mA 0.01 mA + (0.2 % + 2) + (0.2 % + 2) + (1.0 % + 2) + (1.0 % + 2) Burden Voltage (Typical) 2 mV/mA 2 mV/mA Overload 600 mA for 2 minutes maximum Overload Protection 440 mA, 1000 V, FAST fuse AC Conversion Inputs are ac-coupled and calibrated to the rms value of sine wave input. Conversions are true-rms responding and specified from 5 % to 100 % of range. Input signal crest factor can be up to 3 up to 300 mA, decreasing linearly to crest factor