Fluke 1577 FE 1587-1577 Users Manual - Page 20
Table 4. Display Indicators cont., Feature, Description, Table 5. Error Messages, Message, Auto Range - service manual
View all Fluke 1577 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 20 highlights
1587/1577 Users Manual Table 4. Display Indicators (cont.) Feature Auto Range ManualRange 610000mV 2500V 1000V T Description Display range in use Source voltage rating for insulation test: 50, 100, 250, 500 (default) or 1000 V on the 1587. 500 (default) and 1000 V ranges available on the 1577. 50 (default) and 100 V on the 1587T. Insulation test indicator. Appears when insulation test voltage is present. 10 Table 5. Error Messages Message Description batt bat OPEn LEAd 15--Err Appears on the primary display and indicates that the battery is too low for reliable operation. The Meter will not operate at all until the battery is replaced. The b also appears when batt is on the primary display. Appears on the secondary display and indicates that the battery is too low to perform an insulation test. The t button is disabled until the battery is replaced. This message disappears when the rotary switch is turned to any other function. Appears when an open thermocouple is detected. Test lead alert. The message appears briefly and a single beep will sound when you move the switch in or out of the c position. Model detect error. Service Meter if this is displayed.