Fluke 8846A FE 8845A & 8846A Users Manual - Page 66

Measuring Voltage, Caution

Page 66 highlights

8845A/8846A Users Manual Measuring Voltage The Meter is capable of measuring up to 1000 V dc, 750 V ac (8845A) or 1000 V ac (8846A). WCaution To avoid blowing the current fuses and possible damage to other equipment, do not apply voltage to the meter's input until the test leads are properly connected to the input, and the proper voltage function selected. Measuring DC Voltage To make a dc voltage measurement: 1. Press E. The dc voltage icon | will appear to the right of the displayed value as shown below. -0.001698V F1 F2 F3 F4 F5 caw021.eps 2. Connect the test leads to the Meter's inputs as shown in Figure 4-1. 3. Connect the test leads to the circuit and read the measured voltage in the Meter's display. Circuit Under Test INPUT V 2W/4W HI SENSE 4W HI 1000 V CAT I 600V CAT II 300 V LO LO 1V FUSED 100 mA 10 A REAR FRONT caw019.eps Figure 4-1. Input Connections for Voltage, Resistance, and Frequency Measurements. Function modifiers: FILTER An 8 Hertz low-pass filter for rejecting noise and signals above 8 Hz is inserted to stabilize measurements. The filter is active when this soft 4-4

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93

8845A/8846A
Users Manual
4-4
Measuring Voltage
The Meter is capable of measuring up to 1000 V dc, 750 V ac (8845A) or 1000 V ac
(8846A).
W
Caution
To avoid blowing the current fuses and possible damage to
other equipment, do not apply voltage to the meter’s input until
the test leads are properly connected to the input, and the
proper voltage function selected.
Measuring DC Voltage
To make a dc voltage measurement:
1.
Press
E
.
The dc voltage icon
|
will appear to the right of the displayed value as shown
below.
-0.001698
V
F1
F2
F3
F4
F5
caw021.eps
2.
Connect the test leads to the Meter’s inputs as shown in Figure 4-1.
3.
Connect the test leads to the circuit and read the measured voltage in the Meter’s
display.
2
INPUT
SENSE
HI
LO
FRONT
REAR
10A
1V
100
mA
4
W
HI
LO
2W/4W
V
300V
FUSED
1000V CAT I
600V CAT II
Circuit
Under
Test
caw019.eps
Figure 4-1. Input Connections for Voltage, Resistance, and Frequency Measurements.
Function modifiers:
FILTER
An 8 Hertz low-pass filter for rejecting noise and signals above 8 Hz
is inserted to stabilize measurements. The filter is active when this soft