INSTALLATION OF TEMPERATURE RECORDER Installation of recorder MTR-0620LH (1) 1. Attach the fixture to the temperature recorder (MTR-0620LH) by referring to the instruction manual enclosed with the recorder. (Fig. 1) 2. Remove 2 (outside) of 4 screws on the front left of the top panel. Fix the recorder to the top panel by using the 2 screws as shown in Fig. 2. Fig. 1 Fig. 2 3. Remove the caps (both outside and inside) covering the access port for refrigerator and take out the thermal insulation in the port. Pass the recorder sensor to the chamber through the port. Then replace the thermal insulation and caps (both outside and inside). (Fig.3) Note: Make a cut on the caps for access port to pass the capillary tube as shown in the figure below. Fig. 3 4. Fix the recorder sensor to the attachment hole on the back side of the refrigerator by using the large clips provided as an accessory. (Fig. 4) 5. Arrange the capillary tube so that it cannot contact the shelf. Care should be taken for the following points to avoid the crush of the tube when bending the capillary tube. ɾDo not bend the tube of 30 mm from the edge of the sensor. Clip ɾDo not bend and extend the tube repeatedly at the same point. ɾThe bending should be R10ʙR15. Fig. 4 23

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35

23
INSTALLATION OF TEMPERATURE RECORDER
Installation of recorder MTR-0620LH (1)
1.
Attach the fixture to the temperature recorder (MTR-0620LH) by referring to the instruction manual
enclosed with the recorder.
(Fig. 1)
2.
Remove 2 (outside) of 4 screws on the front left of the top panel.
Fix the recorder to the top panel by
using the 2 screws as shown in Fig. 2.
3.
Remove the caps (both outside and inside) covering the access port for refrigerator and take out the
thermal insulation in the port.
Pass the recorder sensor to the chamber through the port.
Then replace
the thermal insulation and caps (both outside and inside).
(Fig.3)
Note:
Make a cut on the caps for access port to
pass the capillary tube as shown in the figure
below.
4.
Fix the recorder sensor to the attachment hole on the back side of the refrigerator by using the large
clips provided as an accessory.
(Fig. 4)
5.
Arrange the capillary tube so that it cannot contact the shelf.
Care should be taken for the following
points to avoid the crush of the tube when bending the capillary tube.
Do not bend the tube of 30 mm from the edge of
the sensor.
Do not bend and extend the tube repeatedly at
the same point.
The bending should be R10
R15.