INSTALLATION OF TEMPERATURE RECORDER 5. Remove the caps (both outside and inside) covering the access port for refrigerator and freezer and take out the thermal insulation in the port. Pass the recorder sensor to the chamber through each port. Then replace the thermal insulation and caps (both outside and inside). Replace the cap on the port for taking out the recorder sensor. (Fig.18) Note: The identification label is adhered to the lead wire to distinguish between the sensor for refrigerator and for freezer. Note: Make a cut on the caps for access port to pass the lead wire as shown in the figure below. Fig. 18 6. Fix the recorder sensor to the attachment hole on the back side of the refrigerator by using the clips enclosed with the recorder fitting (MPR-S7: optional component). (Fig. 9) Clip Fig. 19 7. Fix the recorder sensor to the attachment hole on the left side of the freezer by using the clips enclosed with the recorder fitting (MPR-S7: optional component). (Fig. 20) Fig. 20 29

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35

29
INSTALLATION OF TEMPERATURE RECORDER
5.
Remove the caps (both outside and inside) covering the access port for refrigerator and freezer and
take out the thermal insulation in the port.
Pass the recorder sensor to the chamber through each port.
Then replace the thermal insulation and caps (both outside and inside).
Replace the cap on the port for
taking out the recorder sensor.
(Fig.18)
Note:
The identification label is adhered to the
lead wire to distinguish between the sensor for
refrigerator and for freezer.
Note:
Make a cut on the caps for access port to
pass the lead wire as shown in the figure below.
6.
Fix the recorder sensor to the attachment hole on the back side of the refrigerator by using the clips
enclosed with the recorder fitting (MPR-S7: optional component).
(Fig. 9)
7.
Fix the recorder sensor to the attachment hole on the left side of the freezer by using the clips
enclosed with the recorder fitting (MPR-S7: optional component).
(Fig. 20)
Fig. 18
Fig. 20
Fig. 19
Clip