Sanyo mpr-214f Instruction Manual - Page 26

Care should be taken for the following

Page 26 highlights

INSTALLATION OF TEMPERATURE RECORDER 4. Remove the caps (both outside and inside) covering the access port for refrigerator and take out the thermal insulation in the port. Pass the recorder sensor to the chamber through the port. Then replace the thermal insulation and caps (both outside and inside). Replace the cap on the port for taking out the recorder sensor. (Fig.8) Note: Make a cut on the caps for access port to pass the capillary tube as shown in the figure below. Fig. 8 5. Fix the recorder sensor to the attachment hole on the back side of the refrigerator by using the large clips enclosed with the recorder fitting (MPR-S30: optional component). (Fig. 9) 6. Arrange the capillary tube so that it cannot contact the shelf. Care should be taken for the following points to avoid the crush of the tube when bending the capillary tube. ɾDo not bend the tube of 30 mm from the edge of the sensor. Clip ɾDo not bend and extend the tube repeatedly at the same point. ɾThe bending should be R10ʙR15. Fig. 9 25

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35

25
INSTALLATION OF TEMPERATURE RECORDER
4.
Remove the caps (both outside and inside) covering the access port for refrigerator and take out the
thermal insulation in the port.
Pass the recorder sensor to the chamber through the port.
Then replace
the thermal insulation and caps (both outside and inside).
Replace the cap on the port for taking out the
recorder sensor. (Fig.8)
Note:
Make a cut on the caps for access port to
pass the capillary tube as shown in the figure
below.
5.
Fix the recorder sensor to the attachment hole on the back side of the refrigerator by using the large
clips enclosed with the recorder fitting (MPR-S30: optional component).
(Fig. 9)
6.
Arrange the capillary tube so that it cannot contact the shelf.
Care should be taken for the following
points to avoid the crush of the tube when bending the capillary tube.
Do not bend the tube of 30 mm from the edge of
the sensor.
Do not bend and extend the tube repeatedly at
the same point.
The bending should be R10
R15.
Fig. 8
Fig. 9
Clip