Fluke TiR1 FE Ti10-25-Tir-1 Users Manual - Page 15
Setting the Level (Manual), Setting the Temperature Span (Manual), Level
View all Fluke TiR1 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 15 highlights
Level Thermal Imagers Adjusting the Thermal Image Total Imager Range Span Figure 1. Range and Span Settings exj02.eps Setting the Level (Manual) When put into manual ranging, the level setting adjusts the thermal span within the total temperature range of the Imager. See Figure 1. To set the level: 1. After entering the manual range mode (See the "Setting the Range" section above), press the softkey labeled Level. 2. Press the softkey labeled Up to move the temperature span to higher temperatures or Down to move the span to lower temperatures. 3. Press the softkey labeled Done to set the Imager's range level. The scale along the right side of the display indicates the range. Setting the Temperature Span (Manual) When put into manual ranging, the span setting adjusts the palette over a temperature range within the total range of the Imager. See Figure 1. To adjust the temperature span: 1. After entering the manual range mode (See the "Setting the Range" section above), press the softkey labeled Span. 9