Fluke TiR1 FE Ti10-25-Tir-1 Users Manual - Page 23
Detailed Specifications, Thermal Imagers - tir infrared camera
View all Fluke TiR1 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 23 highlights
Thermal Imagers Detailed Specifications Detailed Specifications Temperature Measurements Temperature Range (not calibrated below -10 °C) TiR and TiR1 20 °C to +100 °C Ti10 20 °C to +250 °C Ti25 20 °C to +350 °C Accuracy TiR and Ti10 5 °C or 5 % (whichever is greater) TiR1 and Ti25 2 °C or 2 % (whichever is greater) Measurement Modes Smooth Auto-Scaling and Manual Scaling On-screen Emissivity Correction Ti25 and TiR1 Only) Imaging Performance Field of View 23 ° x 17 ° Spatial Resolution (IFOV 2.5 mRad Minimum Focus Distance Thermal Lens 15 cm (approx. 6 in) Visible (Visual) Light Lens 46 cm (approx. 18 in) Focus Manual Image Frequency 9 Hz refresh rate Detector Type 160 X 120 Focal Plane Array, uncooled microbolometer Infrared Lens Type 20 mm EFL, F/0.8 lens Thermal Sensitivity (NETD) Ti10 0.2 °C at 30 °C (200 mK) Ti25 0.1 °C at 30 °C (100 mK) TiR 0.1 °C at 30 °C (100 mK) TiR1 0.07 °C at 30 °C (70 mK) Infrared Spectral Band 7.5 μm to 14 μm Visual Camera 640 x 480 resolution Image Presentation Palettes TiR and Ti10 Ironbow, Blue-Red, High Contrast, Grey TiR1 and Ti25 Ironbow, Blue-Red, High Contrast Amber, Hot Metal, Grey Level and Span Smooth Auto-Scaling and Manual Scaling of level and span Minimum Span (in manual mode) 17