Page 23 highlights

Thermal Imagers Detailed Specifications Detailed Specifications Temperature Measurements Temperature Range (not calibrated below -10 °C) TiR and TiR1 20 °C to +100 °C Ti10 20 °C to +250 °C Ti25 20 °C to +350 °C Accuracy TiR and Ti10 5 °C or 5 % (whichever is greater) TiR1 and Ti25 2 °C or 2 % (whichever is greater) Measurement Modes Smooth Auto-Scaling and Manual Scaling On-screen Emissivity Correction Ti25 and TiR1 Only) Imaging Performance Field of View 23 ° x 17 ° Spatial Resolution (IFOV 2.5 mRad Minimum Focus Distance Thermal Lens 15 cm (approx. 6 in) Visible (Visual) Light Lens 46 cm (approx. 18 in) Focus Manual Image Frequency 9 Hz refresh rate Detector Type 160 X 120 Focal Plane Array, uncooled microbolometer Infrared Lens Type 20 mm EFL, F/0.8 lens Thermal Sensitivity (NETD) Ti10 0.2 °C at 30 °C (200 mK) Ti25 0.1 °C at 30 °C (100 mK) TiR 0.1 °C at 30 °C (100 mK) TiR1 0.07 °C at 30 °C (70 mK) Infrared Spectral Band 7.5 μm to 14 μm Visual Camera 640 x 480 resolution Image Presentation Palettes TiR and Ti10 Ironbow, Blue-Red, High Contrast, Grey TiR1 and Ti25 Ironbow, Blue-Red, High Contrast Amber, Hot Metal, Grey Level and Span Smooth Auto-Scaling and Manual Scaling of level and span Minimum Span (in manual mode) 17

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24

Thermal Imagers
Detailed Specifications
17
Detailed Specifications
Temperature Measurements
Temperature Range (not calibrated below -10
°
C)
TiR and TiR1
.................................................
-20
°
C to +100
°
C
Ti10
...............................................................
-20
°
C to +250
°
C
Ti25
...............................................................
-20
°
C to +350
°
C
Accuracy
TiR and Ti10
..................................................
±
5
°
C or 5 % (whichever is greater)
TiR1 and Ti25
................................................
±
2
°
C or 2 % (whichever is greater)
Measurement Modes
.........................................
Smooth Auto-Scaling and Manual
Scaling
On-screen Emissivity Correction
........................
(Ti25 and TiR1 Only)
Imaging Performance
Field of View
......................................................
23
°
x 17
°
Spatial Resolution (IFOV)
..................................
2.5 mRad
Minimum Focus Distance
Thermal Lens
.................................................
15 cm (approx. 6 in)
Visible (Visual) Light Lens
..............................
46 cm (approx. 18 in)
Focus
................................................................
Manual
Image Frequency
...............................................
9 Hz refresh rate
Detector Type
....................................................
160 X 120 Focal Plane Array,
uncooled microbolometer
Infrared Lens Type
.............................................
20 mm EFL, F/0.8 lens
Thermal Sensitivity (NETD)
Ti10
...............................................................
0.2
°
C at 30
°
C (200 mK)
Ti25
...............................................................
0.1
°
C at 30
°
C (100 mK)
TiR
.................................................................
0.1
°
C at 30
°
C (100 mK)
TiR1
...............................................................
0.07
°
C at 30
°
C (70 mK)
Infrared Spectral Band
.......................................
7.5
μ
m to 14
μ
m
Visual Camera
...................................................
640 x 480 resolution
Image Presentation
Palettes
TiR and Ti10
..................................................
Ironbow, Blue-Red, High Contrast,
Grey
TiR1 and Ti25
................................................
Ironbow, Blue-Red, High Contrast
Amber, Hot Metal, Grey
Level and Span
Smooth Auto-Scaling and Manual Scaling of level and span
Minimum Span (in manual mode)