Invacare® HomeFill® Oxygen System 2.3 Provider Information Invacare recommends leaving a full cylinder of oxygen with the patient after setting up the HomeFill compressor. Before leaving a HomeFill Compressor with a user, the following checklist MUST be completed: q Make sure the pressure test on cylinder has not expired. q Instruct the user on the safe operation of the concentrator and review all warnings. q Instruct the user on the safe operation of the HomeFill compressor and review all warnings. q Leave a full cylinder of oxygen with the user. q Leave a copy of all warnings and filling instructions with the user. 2.4 Double-Insulated Products CAUTION! - DO NOT unplug the power cord by pulling on the cord. To unplug, grasp the plug not the cord, otherwise damage may occur. In a double-insulated product, two systems of insulation are provided instead of grounding. No grounding means is required on a double-insulated product, nor is a means for grounding to be added to the product. A double-insulated product is marked with the words "DOUBLE INSULATION" or "DOUBLE INSULATED" or the symbol: 10 2.5 EMC Information WARNING! Risk of Injury or Damage To reduce the risk of injury or product damage from interference with wireless equipment: - Keep the compressor at least 9.8 ft (3.0 m) away from wireless communication equipment such as wireless home network devices, mobile phones, cordless phones and base stations, walkie-talkies, etc. Medical Electrical Equipment needs to be installed and used according to the EMC information in this manual. This equipment has been tested and found to comply with EMC limits specified by IEC/EN 60601-1-2 for Type BF equipment. These limits are determined to provide reasonable protection against electromagnetic interference in a typical home healthcare environment. This equipment is not intended for use in a professional healthcare environment. Portable and mobile RF communications equipment can affect the operation of this equipment. The compressor may be stacked on top on an Invacare HomeFill compatible oxygen concentrator during setup. The compressor operation should be observed to verify normal cylinder filling in either situation. Other devices may experience interference from even the low levels of electromagnetic emissions permitted by the above standard. To determine if the emissions from the compressor are causing the interference, turn the compressor Off. If the interference with the other 1100873-N

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36

Invacare® HomeFill® Oxygen System
2.3
Provider Information
Invacare recommends leaving a full cylinder of
oxygen with the patient after setting up the HomeFill
compressor.
Before leaving a HomeFill Compressor with a user, the
following checklist MUST be completed:
q
Make sure the pressure test on cylinder has not expired.
q
Instruct the user on the safe operation of the
concentrator and review all warnings.
q
Instruct the user on the safe operation of the HomeFill
compressor and review all warnings.
q
Leave a full cylinder of oxygen with the user.
q
Leave a copy of all warnings and filling instructions with
the user.
2.4
Double-Insulated Products
CAUTION!
– DO NOT unplug the power cord by pulling on
the cord. To unplug, grasp the plug not the
cord, otherwise damage may occur.
In a double-insulated product, two systems of insulation
are provided instead of grounding. No grounding means is
required on a double-insulated product, nor is a means for
grounding to be added to the product. A double-insulated
product is marked with the words “DOUBLE INSULATION” or
“DOUBLE INSULATED” or the symbol:
2.5
EMC Information
WARNING!
Risk of Injury or Damage
To reduce the risk of injury or product damage
from interference with wireless equipment:
– Keep the compressor at least 9.8 ft (3.0 m)
away from wireless communication equipment
such as wireless home network devices, mobile
phones, cordless phones and base stations,
walkie-talkies, etc.
Medical Electrical Equipment needs to be installed and used
according to the EMC information in this manual.
This equipment has been tested and found to comply
with EMC limits specified by IEC/EN 60601-1-2 for Type
BF equipment. These limits are determined to provide
reasonable protection against electromagnetic interference
in a typical home healthcare environment. This equipment
is not intended for use in a professional healthcare
environment.
Portable and mobile RF communications equipment can
affect the operation of this equipment. The compressor
may be stacked on top on an Invacare HomeFill compatible
oxygen concentrator during setup. The compressor operation
should be observed to verify normal cylinder filling in either
situation.
Other devices may experience interference from even
the low levels of electromagnetic emissions permitted
by the above standard. To determine if the emissions
from the compressor are causing the interference, turn
the compressor Off. If the interference with the other
10