Invacare® HomeFill® Oxygen System 7 Troubleshooting 7.1 Indicator Lights Indicator Light Color None O2 BELOW NORMAL - YELLOW FILLING - GREEN Compressor Status Effect Compressor is Off. O2 flowing to user if concentrator is on. Compressor is On and warming up (0-3 Cylinder filling not started. minutes after power is On). O2 flowing to user if concentrator is On. O2 from concentrator is < 90% (any time Cylinder filling stops, will restart when after three minute warmup). O2 > 90%. O2 flowing to user if concentrator is On. Compressor is On and cylinder is filling. Cylinder is filling. O2 from concentrator is >90% (any time after three minute warmup). O2 flowing to user if concentrator is On. 26 1100873-N

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36

Invacare® HomeFill® Oxygen System
7
Troubleshooting
7.1
Indicator Lights
Indicator Light Color
Compressor Status
Effect
None
Compressor is Off.
O
2
flowing to user if concentrator is on.
Cylinder filling not started.
Compressor is On and warming up (0-3
minutes after power is On).
O
2
flowing to user if concentrator is On.
Cylinder filling stops, will restart when
O
2
> 90%.
O
2
BELOW NORMAL - YELLOW
O
2
from concentrator is < 90% (any time
after three minute warmup).
O
2
flowing to user if concentrator is On.
Cylinder is filling.
FILLING - GREEN
Compressor is On and cylinder is filling.
O
2
from concentrator is >90% (any time
after three minute warmup).
O
2
flowing to user if concentrator is On.
26
1100873-N