Invacare IOH200 Owners Manual - Page 30
Regulator is On but O
View all Invacare IOH200 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 30 highlights
Invacare® HomeFill® Oxygen System Problem Fill times exceed maximum as recommended fill times OR GREEN filling light stays On. Excessive noise when compressor is On Hissing sound from Regulator when flow selector is Off Regulator is On but O2 is not being delivered 30 Solution 1. Ensure the cylinder regulator is set to Off. 2. Check the recommended fill times shown in the individual cylinder's manual. 3. Check that the cylinder is connected properly. Turn the compressor Off. Remove the cylinder and reinstall. Turn the compressor On. 4. Check the compressor/concentrator line for proper connection and/or damage (i.e. - leaks or kinks). 5. Check that the concentrator liter flow setting is at or below the flow setting listed in 4.5 Turning the Compressor On and Off, page 20. If necessary, adjust the liter flow setting (refer to the concentrator User manual). If the GREEN filling light does not go Off, the compressor and/or concentrator may need service. Contact your home health care provider or Invacare. 1. Call home health care provider or Invacare for service. 1. Turn flow selector Off. 2. If hissing sounds continue, call home health care provider or Invacare. 1. Refill cylinder 2. Check that cannula is straight and not pinched. If damaged, replace. 3. Call home health care provider or Invacare for service. 1100873-N