Invacare IOH200 Owners Manual - Page 18
Pushing DOWN
View all Invacare IOH200 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 18 highlights
Invacare® HomeFill® Oxygen System Fillport covers not shown for clarity. 1. Examine the cylinder gauge. If the cylinder pressure is less than 1500 psig, proceed to STEP 2. For location of the cylinder gauge, refer to the cylinder's manual. If the cylinder pressure is greater than 1500 psig, DO NOT attempt to top off this cylinder. It may not fill. 2. Set the regulator flow dial on the cylinder A to Off. Refer to the cylinder's manual. 18 3. Remove the cylinder and connector fillport covers (if present). 4. Momentarily push DOWN on the outer ring (sleeve) B of the connector fillport C until GREEN dots D are visible to reset the connector. If the outer ring (sleeve) is in the UP position (GREEN dots not visible), the connector fillport will not be able to accept the cylinder fillport E. Pushing DOWN F momentarily will reset the connector fillport (GREEN dots visible) to accept the cylinder fillport. 5. Grasp the cylinder in the area behind the cylinder fillport. 6. Position the cylinder in the compressor cradle G. 7. Align the cylinder fillport with the connector fillport. 8. Pull UP on the outer ring (sleeve) of the connector fillport while pushing DOWN on the cylinder to couple the cylinder fillport into the connector fillport. The cylinder is properly connected when an audible "click" is heard. 1100873-N