Invacare IOH200 Owners Manual - Page 21
c. The FULL GREEN light
View all Invacare IOH200 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 21 highlights
Usage 3. Make sure the cylinder is connected to the compressor. Refer to 4.3 Connecting the Cylinder to the Compressor, page 17. 4. Push the power switch A on the control panel to the On ( I ) position. 5. Examine the control panel B. The following sequence of events should occur: a. 0-3 minutes - The O2 Below Normal (YELLOW) light C will be lit while the compressor warms up. b. After three minutes - The FILLING (GREEN) light D will be on while the cylinder is filling. c. The FULL (GREEN) light E will be on when the cylinder is finished filling. Proceed to STEP 6. The O2 BELOW NORMAL (YELLOW) light may come on if the oxygen level from the concentrator has not yet reached or drops below 90%. The compressor will begin or resume filling when the oxygen level reaches 90%. If light does not go Off within 10 minutes, refer to 7.2 Troubleshooting, page 28. If the ATTENTION (RED) light and audible alarm F are on refer to 7.2 Troubleshooting, page 28. 6. Remove the full cylinder. Refer to 4.4 Disconnecting the Cylinder from the Compressor, page 19. 7. Push the power switch to the Off (O) position. 1100873-N 21