Invacare IOH200 Owners Manual - Page 29
Turning the Compressor On and, Off,
View all Invacare IOH200 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 29 highlights
Problem Compressor cycles on and Off or O2 Below Normal (yellow) and filling (green) lights are alternating or O2 below normal (yellow) light goes On and Off O2 below normal (YELLOW) light stays On Troubleshooting Solution 1. Check that the concentrator liter flow setting is at or below the flow setting listed in 4.5 Turning the Compressor On and Off, page 20. If necessary, adjust the liter flow setting (refer to the concentrator User manual). Wait at least 20 minutes for the GREEN light to appear. If it does not appear, proceed to step 2. 2. The concentrator may need service, contact your home health care provider or Invacare. 1. Ensure the concentrator has warmed up for at least 20 minutes. If it has not been 20 minutes, turn the compressor Off until the concentrator warms up. If the YELLOW light still remains on proceed to step 2. 2. Inspect the connection between the compressor and the concentrator for damage or pinching. • If damaged, replace. • If pinched, straighten tubing. Turn the compressor Off for at least 10 seconds. Turn the compressor On. If the YELLOW light still remains on after 10 minutes, proceed to step 3. 3. The concentrator may need service, contact your home health care provider or Invacare. 1100873-N 29