Problem Compressor cycles on and Off or O2 Below Normal (yellow) and filling (green) lights are alternating or O2 below normal (yellow) light goes On and Off O2 below normal (YELLOW) light stays On Troubleshooting Solution 1. Check that the concentrator liter flow setting is at or below the flow setting listed in 4.5 Turning the Compressor On and Off, page 20. If necessary, adjust the liter flow setting (refer to the concentrator User manual). Wait at least 20 minutes for the GREEN light to appear. If it does not appear, proceed to step 2. 2. The concentrator may need service, contact your home health care provider or Invacare. 1. Ensure the concentrator has warmed up for at least 20 minutes. If it has not been 20 minutes, turn the compressor Off until the concentrator warms up. If the YELLOW light still remains on proceed to step 2. 2. Inspect the connection between the compressor and the concentrator for damage or pinching. • If damaged, replace. • If pinched, straighten tubing. Turn the compressor Off for at least 10 seconds. Turn the compressor On. If the YELLOW light still remains on after 10 minutes, proceed to step 3. 3. The concentrator may need service, contact your home health care provider or Invacare. 1100873-N 29

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36

Troubleshooting
Problem
Solution
Compressor cycles on and Off
or
O
2
Below Normal
(yellow) and filling (green) lights are alternating
or
O
2
below normal (yellow) light goes On and Off
1. Check that the concentrator liter flow setting is at or below
the flow setting listed in
4.5 Turning the Compressor On and
Off, page 20
.
If necessary, adjust the liter flow setting (refer to the
concentrator User manual). Wait at least 20 minutes for the
GREEN light to appear. If it does not appear, proceed to
step 2.
2. The concentrator may need service, contact your home
health care provider or Invacare.
O
2
below normal (YELLOW) light stays On
1. Ensure the concentrator has warmed up for at least 20
minutes. If it has not been 20 minutes, turn the compressor
Off until the concentrator warms up. If the YELLOW light still
remains on proceed to step 2.
2. Inspect the connection between the compressor and the
concentrator for damage or pinching.
• If damaged, replace.
• If pinched, straighten tubing. Turn the compressor Off for
at least 10 seconds. Turn the compressor On. If the YELLOW
light still remains on after 10 minutes, proceed to step 3.
3. The concentrator may need service, contact your home
health care provider or Invacare.
1100873-N