Invacare XPO100 Owners Manual - Page 15
Typical Product Parameters - oxygen concentrator
View all Invacare XPO100 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 15 highlights
SECTION 5-TYPICAL PRODUCT PARAMETERS Oxygen Concentration:* *Based on an atmospheric pressure of 14.7 psi (101 kPa) at 70°F (21°C) 87% to 95.6%, after initial warm-up period (approximately 5 minutes) Oxygen Flow: Pulse flow delivery. Bolus Volume Ranges From 300 - 840cc settings 1 to 5. Dimensions: 10" high x 7" wide x 4" deep (25.4 cm high x 17.8 cm wide x 10.2 cm wide) Weight: 6.0 lbs (2.72 kg); 1.3 lbs (0.68 kg) for supplemental battery Battery Duration (times are approximate): Setting 1 = 3.5 hrs Setting 2 = 2.5 hrs Setting 3 = 2.0 hrs Setting 4 = 1.5 hrs Setting 5 = 1.0 hrs Battery Recharge Time: NOTE: Recharge time increases if battery is charging while unit is running. 4 hours Humidity Operating Humidity: 15% to 60% non condensing Storage Humidity: up to 95% non condensing Temperature Range: (All power sources) Operating temperature: 41°F to 95°F (5°C to 35°C) Storage temperature: -2°F to 140°F (-20°C to 60°C) Extended Temperature Range: (Using AC or DC adapters) Operating Temperature: 95°F to 104° (35°C to 40°C) Continuous use - all settings Extended Temperature Range (Using Internal Battery) Operating Temperature: 95°F to 104° (35°C to 40°C) Continuous use - settings 1, 2 and 3 45 minutes (max) - setting 4 30 minutes (max) - setting 5 Part No 1148112 15 Invacare XPO2™