Invacare XPO100 Owners Manual - Page 9
Maintenance, Radio Frequency Interference
View all Invacare XPO100 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 9 highlights
SECTION 1-GENERAL GUIDELINES For optimum performance, Invacare recommends that each concentrator be on and running for a minimum of 5 minutes. Shorter periods of operation may reduce maximum product life. The XPO2 is not designed to be used with a humidifier. Use of this device with a humidifier may impair performance and/or damage the equipment. When your automobile is turned Off, disconnect the car accessory power supply and remove the XPO2 from the automobile. NEVER allow the XPO2 to be stored in a very hot or cold automobile or in other similar, high or low, temperature environments. Refer to Typical Product Parameters on page 14. DO NOT operate in temperatures below 41° F (5°C) or above 104° F (40° C) for extended periods of time. The XPO2 MUST be used in an upright position. Invacare recommends the XPO2 not be used in the rain. Maintenance The XPO2 was specifically designed to minimize routine preventive maintenance. Only professionals of the healthcare field or persons fully conversant with this process such as factory trained personnel should perform preventive maintenance or performance adjustments on the oxygen concentrator. Radio Frequency Interference This equipment has been tested and found to comply with EMC limits specified by IEC/EN 60601‐1‐2. These limits are designed to provide a reasonable protection against electromagnetic interference in a typical medical installation. Part No 1148112 9 Invacare XPO2™