SECTION 8-TROUBLESHOOTING GUIDE SECTION 8-TROUBLESHOOTING GUIDE If your XPO2 Portable Oxygen Concentrator fails to operate properly, refer to the following chart for possible causes and solutions. If necessary, contact your service provider. ALARM/WARNING: SYMPTOM PROBABLE CAUSE START-UP INDICATOR XPO2 does not operate when On/Off button was not held down long enough. On/Off button pressed. OR Front panel lights DO NOT illuminate upon start-up OR Internal battery is discharged (and supplemental battery if connected to the unit). Malfunction All front panel lights go dark. NO BREATH DETECTED ALARM: Continuous audible beep and red alarm indicator is on XPO2 has not detected a breath for a predetermined time period. continuously (not blinking) upon start up. SOLUTION Try to power up the unit again while continuing to press the On/Off button until the front panel lights begin to illuminate. This takes typically 2-3 seconds. Connect XPO2 to either the AC or DC charger and retry. Change to an alternate oxygen supply and contact your equipment provider. 1. Verify the cannula is connected, not kinked, properly positioned and you are breathing through your nose. 2. If the alarm continues, change to another source of oxygen and contact your equipment provider. Invacare XPO2™ 40 Part No 1148112

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48

SECTION 8—TROUBLESHOOTING GUIDE
Invacare XPO
2
40
Part No 1148112
SECTION 8—TROUBLESHOOTING
GUIDE
If your XPO
2
Portable Oxygen Concentrator fails to operate
properly, refer to the following chart for possible causes and
solutions. If necessary, contact your service provider.
ALARM/WARNING:
SYMPTOM
PROBABLE CAUSE
SOLUTION
START-UP INDICATOR
XPO
2
does not
operate when
On/Off button
pressed.
OR
Front panel lights
DO NOT
illuminate upon
start-up
OR
All front panel
lights go dark.
On/Off button was not held
down long enough.
Try to power up the
unit again while
continuing to press the
On/Off button until the
front panel lights begin
to illuminate. This takes
typically 2-3 seconds.
Internal battery is
discharged (and
supplemental battery if
connected to the unit).
Connect XPO
2
to either
the AC or DC charger
and retry.
Malfunction
Change to an alternate
oxygen supply and
contact your equipment
provider.
NO BREATH DETECTED ALARM:
Continuous
audible beep and
red alarm
indicator is on
continuously (not
blinking) upon
start up.
XPO
2
has not detected a
breath for a predetermined
time period.
1. Verify the cannula is
connected, not kinked,
properly positioned and
you are breathing
through your nose.
2. If the alarm continues,
change to another
source of oxygen and
contact your equipment
provider.