SECTION 8-TROUBLESHOOTING GUIDE ALARM/WARNING: SYMPTOM PROBABLE CAUSE CHARGING INDICATOR With the unit Power source is no good, or turned Off and there is a loose connection. with the AC or DC charging adapter plugged into the unit, no Battery Gauge segments are Internal Battery is outside the allowed temperature range for charging. illuminated. Battery Gauge is dark. LOW BATTERY WARNING: There is a single Internal battery requires audible beep and charging. the green 25% battery indicator is blinking. LOW BATTERY ALARM: There is a double Internal battery requires audible beep and charging. the green 25% battery indicator is rapidly blinking. BATTERY DISCHARGED ALARM: There is a triple Internal battery is audible beep and completely discharged and the green 25% needs to be recharged. battery indicator is blinking very rapidly. SOLUTION Try another power outlet and check connections to charger and unit. Allow unit to cool down to less than 95ºF, or warm up to 50ºF. Connect XPO2 to its AC or DC power adapter, or to its supplemental battery pack. 1. Connect XPO2 to its AC or DC power adapter, or to its supplemental battery pack. 2. If no other power source is available, change to a reserve oxygen supply. 1. Connect XPO2 to its AC or DC power adapter, or to its supplemental battery pack. 2. If no other power source is available, change to a reserve oxygen supply. Part No 1148112 41 Invacare XPO2™

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48

SECTION 8—TROUBLESHOOTING GUIDE
Part No 1148112
41
Invacare XPO
2
CHARGING INDICATOR
With the unit
turned Off and
with the AC or
DC charging
adapter plugged
into the unit, no
Battery Gauge
segments are
illuminated.
Battery Gauge is
dark.
Power source is no good, or
there is a loose connection.
Try another power
outlet and check
connections to charger
and unit.
Internal Battery is outside
the allowed temperature
range for charging.
Allow unit to cool down
to less than 95ºF, or
warm up to 50ºF.
LOW BATTERY WARNING:
There is a single
audible beep and
the green 25%
battery indicator is
blinking.
Internal battery requires
charging.
Connect XPO
2
to its AC
or DC power adapter,
or to its supplemental
battery pack.
LOW BATTERY ALARM:
There is a double
audible beep and
the green 25%
battery indicator is
rapidly blinking.
Internal battery requires
charging.
1. Connect XPO
2
to its
AC or DC power
adapter, or to its
supplemental battery
pack.
2. If no other power
source is available,
change to a reserve
oxygen supply.
BATTERY DISCHARGED ALARM:
There is a triple
audible beep and
the green 25%
battery indicator is
blinking very
rapidly.
Internal battery is
completely discharged and
needs to be recharged.
1. Connect XPO
2
to its
AC or DC power
adapter, or to its
supplemental battery
pack.
2. If no other power
source is available,
change to a reserve
oxygen supply.
ALARM/WARNING:
SYMPTOM
PROBABLE CAUSE
SOLUTION