SECTION 8-TROUBLESHOOTING GUIDE ALARM/WARNING: SYMPTOM PROBABLE CAUSE OPERATING ALARM: Continuous Abnormal operation has audible beep and been detected. red alarm indicator is illuminated. AND SOLUTION 1. Turn Off the unit. 2. Turn On the unit. 3. If the alarm continues, change to another source of oxygen and contact your equipment provider Flow indicators 3 & 4 are illuminated. COMPRESSOR ALARM: Continuous Abnormal compressor audible beep and operation has been red alarm detected. indicator is illuminated. AND 1. Turn Off the unit. 2. Turn On the unit. 3. If the alarm continues, change to another source of oxygen and contact your equipment provider Flow indicators 4 & 5 are illuminated. SYSTEM ALARM: Continuous Abnormal system operation audible beep and has been detected. red alarm indicator is illuminated. AND 1. Turn Off the unit. 2. Turn On the unit. 3. If the alarm continues, change to another source of oxygen and contact your equipment provider Flow indicators 3, 4 & 5 are illuminated. Invacare XPO2™ 44 Part No 1148112

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48

SECTION 8—TROUBLESHOOTING GUIDE
Invacare XPO
2
â„¢
44
Part No 1148112
OPERATING ALARM:
Continuous
audible beep and
red alarm
indicator is
illuminated.
AND
Flow indicators
3 & 4 are
illuminated.
Abnormal operation has
been detected.
1. Turn Off the unit.
2. Turn On the unit.
3. If the alarm continues,
change to another
source of oxygen and
contact your equipment
provider
COMPRESSOR ALARM:
Continuous
audible beep and
red alarm
indicator is
illuminated.
AND
Flow indicators
4 & 5 are
illuminated.
Abnormal compressor
operation has been
detected.
1. Turn Off the unit.
2. Turn On the unit.
3. If the alarm continues,
change to another
source of oxygen and
contact your equipment
provider
SYSTEM ALARM:
Continuous
audible beep and
red alarm
indicator is
illuminated.
AND
Flow indicators
3, 4 & 5 are
illuminated.
Abnormal system operation
has been detected.
1. Turn Off the unit.
2. Turn On the unit.
3. If the alarm continues,
change to another
source of oxygen and
contact your equipment
provider
ALARM/WARNING:
SYMPTOM
PROBABLE CAUSE
SOLUTION