Invacare XPO100 Owners Manual - Page 31
Adjusting the Pulse Flow Setting
View all Invacare XPO100 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 31 highlights
Oxygen Outlet Port SECTION 6-OPERATING INSTRUCTIONS Nasal Cannula DETAIL "B" DETAIL "A" FIGURE 6.8 Connecting/Positioning the Nasal Cannula Adjusting the Pulse Flow Setting CAUTION It is very important to select the prescribed level of oxygen flow. DO NOT increase or decrease the flow unless a change has been prescribed by your physician or therapist. NOTE: For this procedure, refer to FIGURE 6.9 on page 32. 1. Turn On the XPO2 by pressing and holding the power On/Off button for approximately one second. Each time the XPO2 is turned On, a brief alarm will sound. The XPO2 pulse flow setting at power‐up will be the same flow setting selected the last time the unit was turned Off. 2. The XPO2 has five pulse flow settings. Press the flow selection (+ or ‐) button until your prescribed flow setting (1 to 5) is illuminated by a blue indicator light. Part No 1148112 31 Invacare XPO2™