Using the Color Management 11 Calibrating a single display 1. Click to select the display box that represents the display you want to calibrate on the Display Wall tab, and then click Calibrate Single. • In the Color Management, each display is represented by a display box on the Display Wall tab. For more information about the display box, refer to "Display Wall tab" on page 18. • If the display is connected to the computer successfully, the "connected" icon appears on the upper right corner of the display box. If the "disconnected" icon appears, refer to "Troubleshooting" on page 23 for how to solve the connection problem. • The power-on time of the display must be longer than 60 minutes before calibration. Otherwise a dialog asking you warm-up the display will appear after you click the Calibrate Single button. 2. The calibration window appears. Click Run to proceed. Display box You can also click Track to measure the CIE1931 (x,y) values of current W,R,G and B colors at level 255 for further analysis.

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24

Using the Color Management
11
Calibrating a single display
1.
Click to select the display box that represents the display you want to calibrate on
the
Display Wall
tab, and then click
Calibrate Single
.
2.
The calibration window appears. Click
Run
to proceed.
• In the Color Management, each display is represented by a display box
on the
Display Wall
tab. For more information about the display box,
refer to
"Display Wall tab" on page 18
.
• If the display is connected to the computer successfully, the “connected”
icon
appears on the upper right corner of the display box. If the
“disconnected” icon
appears, refer to
"Troubleshooting" on page 23
for how to solve the connection problem.
• The power-on time of the display must be longer than 60 minutes before
calibration. Otherwise a dialog asking you warm-up the display will
appear after you click the
Calibrate Single
button.
You can also click
Track
to measure the CIE1931 (x,y) values of current W,R,G and B colors at
level 255 for further analysis.
Display box