Canon DP-V2420 User Manual - Page 27

Press the jog dial and select Start., Finish calibration.

Page 27 highlights

3. Press the jog dial and select "Start". Please follow the information indicated on the screen. 4. Initialize the sensor. Set the mode dial of the Universal Measuring Probe to "0-CAL". Press the jog dial of the video display, select "OK", and execute initialization. 5. Place the universal measuring probe pointing at the center of the video display. Set the mode dial of the universal measuring probe to "MEAS" and place the probe as shown below according to the displayed content. Press the jog dial of the video display, select "OK", and execute calibration. Main Unit Konica Minolta Universal Measuring Probe CA-310 support: CA-PU32, CA-PU35 CA-210 support: CA-PU12, CA-PU15 Konica Minolta Display Color Analyzer CA-310, CA-210 6. Finish calibration. When the message "Calibration is completed". is displayed, press the jog dial and select "OK". • If the message "Calibration error." is displayed. Calibration has been terminated due to an error. The main unit returns to the state before calibration. ( 95) • To cancel calibration Press the jog dial during calibration and select "Cancel". The main unit returns to the state before calibration. Note • Due to the characteristic of LCD panel and individual difference of CA-310 and CA-210, the calibration results may differ. • Perform matrix calibration of the display color analyzer prior to calibration. If calibration is performed without performing matrix calibration, an error may occur. Refer to the CA-310 and CA-210 instruction manual for the detail operation. • Warming-up is necessary to stabilize the brightness of the video display. Wait at least 10 minutes after turning on the power before calibration. • Perform calibration in a dark room so that no external light enters the sensor. If external light enters the sensor, low brightness characteristics cannot be calibrated correctly. Operating the Video Display 27

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108

27
Operating the Video Display
3. Press the jog dial and select "Start".
Please follow the information indicated on the screen.
4. Initialize the sensor.
Set the mode dial of the Universal Measuring Probe to "0-CAL".
Press the jog dial of the video display, select "OK", and execute initialization.
5. Place the universal measuring probe pointing at the center of the video display.
Set the mode dial of the universal measuring probe to "MEAS" and place the probe as shown below according to
the displayed content. Press the jog dial of the video display, select "OK", and execute calibration.
Konica Minolta Universal Measuring Probe
CA-310 support: CA-PU32, CA-PU35
CA-210 support: CA-PU12, CA-PU15
Konica Minolta
Display Color Analyzer CA-310, CA-210
Main Unit
6. Finish calibration.
When the message "Calibration is completed". is displayed, press the jog dial and select "OK".
• If the message "Calibration error." is displayed.
Calibration has been terminated due to an error. The main unit returns to the state before calibration. (
95)
• To cancel calibration
Press the jog dial during calibration and select "Cancel". The main unit returns to the state before calibration.
Note
• Due to the characteristic of LCD panel and individual difference of CA-310 and CA-210, the calibration results may differ.
• Perform matrix calibration of the display color analyzer prior to calibration. If calibration is performed without performing
matrix calibration, an error may occur. Refer to the CA-310 and CA-210 instruction manual for the detail operation.
• Warming-up is necessary to stabilize the brightness of the video display. Wait at least 10 minutes after turning on the power
before calibration.
• Perform calibration in a dark room so that no external light enters the sensor. If external light enters the sensor, low
brightness characteristics cannot be calibrated correctly.