Fluke 1735 User Manual - Page 38
Split Phase
View all Fluke 1735 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 38 highlights
1735 Users Manual Split Phase For Split Phase, the Neutral is center-tapped and there are two hot legs which correspond to A and B test Leads. AB is the voltage from phase to phase, which is twice that of each individual hot leg. Refer to Figure 7 and connect the leads as follows: Voltage: Network Test Leads Line 1 Line 2 Line 1 N A (L1) B (L2) C (L3) N Current: Network A (L1) B (L2) Line 1 Not connected Neutral N A (L1) B (L2) C(L3) N Test Leads 28
1735
Users Manual
28
Split Phase
For Split Phase, the Neutral is center-tapped and there are two hot legs which
correspond to A and B test Leads. AB is the voltage from phase to phase,
which is twice that of each individual hot leg. Refer to Figure 7 and connect
the leads as follows:
Voltage:
Network
Test Leads
Line 1
A (L1)
Line 2
B (L2)
Line 1
C (L3)
N
N
Current:
Network
Test Leads
A (L1)
A (L1)
B (L2) Line 1
B (L2)
Not connected Neutral
C(L3)
N
N