Fluke 1735 User Manual - Page 51
Save, Logged Events, Power Log, Hold/Run, Save/Enter, Fluke Power Log - data logger software
View all Fluke 1735 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 51 highlights
Power Logger Measuring Functions MAX values of the half-cycle RMS values are shown as graphs. The screenshots recorded by this method are saved as individual pictures and can be viewed later, or the data can be displayed with the Power Log software. A total of 999 events can be recorded. In the LC display the phase and the number of recordings are displayed. edx029.bmp 21 Switch between the individual events (if there is more than one). This is also possible if the recording has been stopped and you want to evaluate the stored events. With Hold/Run you can stop/start the measurement or you can also start a new measurement. Save With Save/Enter you will get a screenshot and thus store the actual display in the memory location shown. Logged Events You can download logged events using Fluke Power Log software package. The Fluke Power Log software presents the event data in a variety of formats: • graphs similar to those displayed on the instrument 41