Fluke 1735 User Manual - Page 68
EMC, Safety, V-RMS Wye Measurement, coefficient below. - battery pack
![]() |
View all Fluke 1735 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 68 highlights
1735 Users Manual Note The above terms are defined in European Standards. To calculate the specification at any point in the working temperature range, use the temperature coefficient below. Temperature coefficient: ±0.1 % of the measured value per K. Intrinsic error: Refers to reference temperature, max. deviation is guaranteed for 2 years. Operating error: Refers to operating temperature range, max. deviation is guaranteed for 2 years. Climatic class: C1 (IEC 654-1) -5 °C to +45 °C, 5 % to 95 % RH, no dew Housing: Cycoloy shock and scratch proof thermoplast V0-type (nonflammable) with rubber protection holster EMC Emission: Immunity: Power supply: Operation time with battery: Dimensions: Weight: IEC 61326-1:2006 class B IEC 61326-1:2006 NiMH battery-pack, with ac adapter (15 V to 20 V/0.8 A) Typical >8 h with bright backlight, >10 h with low backlight, and 24 h without backlight 240 x 180 x 110 mm (6.1 x 4.6 x 2.8 in) 1.7 kg (3.75 lb), including battery Safety Safety: EN/IEC 61010-1:2001 (2nd ed.) 600 V CAT III, double or enforced insulation Pollution Degree: 2 Protection: IP65; EN60529 (refers only to the main housing without the battery compartment) RMS values are measured with a 20 ms resolution. V-RMS Wye Measurement Measuring range: Intrinsic error: Operating error: Resolution: 57 / 66 / 110 / 120 / 127 / 220 / 230/ 240 / 260 / 277 / 347 / 380 / 400 / 417 / 480 V AC ±(0.2 % of measured value. + 5 digits) ±(0.5 % of m. v. + 10 digit) 0.1 V 58