Fluke 1735 User Manual - Page 71
Harmonics, Events, Accuracy
View all Fluke 1735 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 71 highlights
Harmonics Measuring range: Power Logger Specifications 1 to 50th harmonic (
Power Logger
Specifications
61
Harmonics
Measuring range:
1 to 50
th
harmonic (<50 % von Vm)
Accuracy:
Vm, Im, THDV, THDI:
According IEC 1000-4-7, class B
Vm, Im, THDV, THDI:
According IEC 1000-4-7, class B
Vm
≥
3 % Vn:
5 % V
Vm < 3 % Vn:
0.15 % Vn
Im
≥
10 % In:
5 % Im
Im < 10 % In:
0.5 % In
THDV:
for THD <3 % - <0.15 % at Vn
for THD
≥
3 % - <5
% at Vn
THDI:
for THD <10 % - <0.5 % at In
for THD
≥
10 % - <5 % at In
Events
Detection of voltage dips, voltage swells and voltage interruptions with a 10 ms resolution
and measuring error of the half period sine wave of RMS.
Intrinsic error:
±
(1 % of m. v. + 10 digit)
Operating error:
±
(2 % of m. v. + 10 digit)
Resolution:
0.1 V