Hayward TriStar® VS Owners Manual - Page 10

TriStar Variable Speed Pump Control, USE ONLY HAYWARD GENUINE REPLACEMENT PARTS

Page 10 highlights

TriStar Variable Speed Pump Control ____ Page 10 of 16 Press 5 (Next) to define the Timer-On time. To change these settings, confirm by pressing 6 (OK). An underlining cursor will appear below hours value. Press 5 (Next) or 2 (Prev) to select hours or minutes to be set, and use Time Up/Dn buttons to select desired value. Press 6 (OK) to store this option. *TMR.ON 8:00am *TMR.ON 8:00am *TMR.ON 8:00am *TIMER SAVED! Press 5 (Next) to define the Timer-Off time. This setting is changed the same as above. *TMR.OFF 5:00pm SETUP mode: PRIME HEIGHT function This menu allows the user to view the prime mode that was selected during initial setup. Press 6 (OK) to display the prime mode that was selected during initial setup. After initial setup, this becomes a read-only display. *MORE THAN 5 FT (for example) SETUP mode: PRESET SETUP function This menu allows the user to define the pump speed and duration for each preset, which should be customized for each installation. Use 5 (Next) or 2 (Prev) to select preset to be setup. For example, user stops on mode 4 (Medium High). Using % Flow Up/Dn buttons for currently displayed mode number causes desired pump speed to be set within predefined range. Using Time Up/Dn buttons for currently displayed mode number causes desired time to be set within predefined range. *SELECT PRESET: 4 *P4.SPEED= 20% *P4.TIME= 1h 30m Press 6 (OK) to store preset time *P4.STORED! and % flow settings. Each preset is set the same as above. SETUP mode: SEQUENCER function This menu allows the For example, user stops on *S1:P1 0h30min user to define the sequence step 1 (S1). active preset and duration for each Use % Flow Up/Dn to select *S1:P3 1h45min sequence step, which desired active preset for displayed *S1:STOP 3h30min should be customized sequence step, or to select Stop *S1:EMPTY for each installation. Mode or Empty. If Empty is Use 5 (Next) or 2 selected, the sequence step will (Prev) to select be skipped, and the next step will sequence step to be be active. When presets or Stop setup. are selected for the sequence step, the step time defaults to the time Empty may be set for the displayed preset. This selected if the user step time may be changed using does not need all of the Time Up/Dn buttons. the sequence steps. Stop may be selected Press 6 (OK) to store sequence *S1.STORED! if the user wishes to step settings. Each sequence step stop the pump during is set the same as above. peak energy periods or another time of day. WWW.HAYWARDPOOL.COM USE ONLY HAYWARD GENUINE REPLACEMENT PARTS

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16

TriStar Variable Speed Pump Control
____
____________________________Page 10 of 16
WWW.HAYWARDPOOL.COM
USE ONLY HAYWARD GENUINE REPLACEMENT PARTS
Press 5 (Next) to define the
Timer-On time.
To change these
settings, confirm by pressing 6
(OK).
An underlining cursor will
appear below hours value.
Press
5 (Next) or 2 (Prev) to select
hours or minutes to be set, and
use Time Up/Dn buttons to select
desired value.
Press 6 (OK) to
store this option.
Press 5 (Next) to define the
Timer-Off time.
This setting is
changed the same as above.
*TMR.ON
8:00am
*TMR.ON
8
:00am
*TMR.ON
8:00
am
*TIMER SAVED!
*TMR.OFF
5:00pm
SETUP mode:
PRIME HEIGHT
function
This menu allows the
user to view the prime
mode that was
selected during initial
setup.
Press 6 (OK) to display the prime
mode that was selected during
initial setup.
After initial setup,
this becomes a read-only display.
*MORE THAN 5 FT
(for example)
SETUP mode:
PRESET SETUP
function
This menu allows the
user to define the
pump speed and
duration for each
preset, which should
be customized for
each installation.
Use
5 (Next) or 2 (Prev) to
select preset to be
setup.
For example, user stops on mode
4 (Medium High).
Using % Flow Up/Dn buttons for
currently displayed mode number
causes desired pump speed to be
set within predefined range.
Using Time Up/Dn buttons for
currently displayed mode number
causes desired time to be set
within predefined range.
Press 6 (OK) to store preset time
and % flow settings.
Each preset
is set the same as above.
*SELECT PRESET: 4
*P4.SPEED=
20%
*P4.TIME=
1h 30m
*P4.STORED!
SETUP mode:
SEQUENCER
function
This menu allows the
user to define the
active preset and
duration for each
sequence step, which
should be customized
for each installation.
Use 5 (Next) or 2
(Prev) to select
sequence step to be
setup.
Empty may be
selected if the user
does not need all of
the sequence steps.
Stop may be selected
if the user wishes to
stop the pump during
peak energy periods or
another time of day.
For example, user stops on
sequence step 1 (S1).
Use % Flow Up/Dn to select
desired active preset for displayed
sequence step, or to select Stop
Mode or Empty.
If Empty is
selected, the sequence step will
be skipped, and the next step will
be active.
When presets or Stop
are selected for the sequence step,
the step time defaults to the time
set for the displayed preset.
This
step time may be changed using
the Time Up/Dn buttons.
Press 6 (OK) to store sequence
step settings.
Each sequence step
is set the same as above.
*S1:P1
0h30min
*S1:P3
1h45min
*S1:STOP
3h30min
*S1:EMPTY
*S1.STORED!