Fluke 116 CAL Product Manual

Fluke 116 CAL Manual

Fluke 116 CAL manual content summary:

  • Fluke 116 CAL | Product Manual - Page 1
    110/113/114/115/116/117 True-rms Multimeter Calibration Information September 2006 Rev. 2, 5/20 ©2006-2020 Fluke Corporation. All rights reserved. All product names are trademarks of their respective companies. MyFlukeStore www. Specifications are subject to change without notice. Shop for
  • Fluke 116 CAL | Product Manual - Page 2
    Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price. Fluke Fluke authorized service center within the warranty period. To obtain warranty service, contact your nearest Fluke authorized service
  • Fluke 116 CAL | Product Manual - Page 3
    7 113 ...7 Backlight Test 7 Keypad Test...7 110/114/115/116/117 7 113 ...7 Preparing for the Performance Tests 8 Testing Temperature (116 only 8 Testing the VoltAlert Function (117 only 12 Calibration Adjustment 14 Calibration Adjustment Password 14 Changing the Password 14 Restoring the
  • Fluke 116 CAL | Product Manual - Page 4
    verify the performance of the Fluke Models 110, 113, 114, 115, 115C, 116, 116C, 117, and 117C True RMS Multimeter (the Meter or Product). When specific models are noted in this manual, the "C" version is also included. For example, when the listed model is 115, the instructions are applicable to the
  • Fluke 116 CAL | Product Manual - Page 5
    110/113/114/115/116/117 Calibration Information Maintenance Maintenance of the Meter consists of battery and fuse If the display shows any other value, have the Meter serviced. See How to Contact Fluke earlier in this document. Figure 1. Test the Fuse
  • Fluke 116 CAL | Product Manual - Page 6
    This notice explains how to minimize damage to these components. 1. Understand the problem. 2. Learn the guidelines for proper handling. 3. Use the proper procedures, tip soldering irons only. static-free work station. MyFlukeStore Shop for Fluke products online at: www. 3 .com 1.888.610.7664
  • Fluke 116 CAL | Product Manual - Page 7
    110/113/114/115/116/117 Calibration Information See Figure 2 for disassembly. Figure 2. Disassembly To remove it with an 11 A, 1000 V, FAST fuse having a minimum interrupt rating of 17 000 A. Use only Fluke PN 803293. 6. To re-assemble the Meter, first attach the case bottom to the case top, then
  • Fluke 116 CAL | Product Manual - Page 8
    of the features for each Meter. Table 1. Features Switch Position Measurement Function 110 113 114 115 116 117 OFF The Meter is turned off.     x Automatically selects ac or dc and AC amps are dc-coupled. MyFlukeStore Shop for Fluke products online at: www. 5 .com 1.888.610.7664
  • Fluke 116 CAL | Product Manual - Page 9
    116/117 Calibration Requirements Equipment 5522 Multi-product Calibrator (or equivalent) Fluke 80 AK K-type Thermocouple Capacitance Temperature Frequency Temperature Temperature VoltAlert [1] For accuracy specifications, see the Users Manual for the product. Range[1] 10 mV to 600 V 600 μA to 10
  • Fluke 116 CAL | Product Manual - Page 10
    test the backlight, press Q and verify that the backlight turns on. Keypad Test 110/114/115/116/117 To test the keypad: 1. Turn the rotary switch to . 2. Push each button separately. then back to any function position. MyFlukeStore Shop for Fluke products online at: www. 7 .com 1.888.610.7664
  • Fluke 116 CAL | Product Manual - Page 11
    115, 116, 117) or Table 4 (113). If the display reading falls outside of the range shown, the Meter requires calibration adjustment or repair. Testing Temperature (116 only) Connect 10.52 8. 40.00 M 30 MΩ 28.48 to 31.52 MyFlukeStore 8 Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 116 CAL | Product Manual - Page 12
    Performance Tests Table 3. DMM Performance Tests: 110/114/115/116/117 (cont.) Step Function Range Applied Display Reading[1] 110 114 115 116 117 9. R 10. Continuity 600 Ω 600 Ω 20 NA 7.18 to 10.82 MyFlukeStore Shop for Fluke products online at: www. 9 .com 1.888.610.7664
  • Fluke 116 CAL | Product Manual - Page 13
    : 110/114/115/116/117 (cont.) Display Reading[1] Step Function Range Applied 110 114 115 116 117 Set calibrator to standby, reconfigure using a Fluke 9100 calibrator, the Calibrator Frequency mode must be used to obtain accurate frequency. [2] To keep from tripping the calibrator to standby
  • Fluke 116 CAL | Product Manual - Page 14
    1 kHz 575.7 to 624.3 21.  Capacitance 22. 1000 nF 9999 μF Open 900 μF 0 to 2 881 to 919 [1] Manually select the range by pressing q. [2] To keep from tripping the calibrator to standby, ramp up the voltage in 50 V increments with a five-second delay between increments. MyFlukeStore Shop for
  • Fluke 116 CAL | Product Manual - Page 15
    110/113/114/115/116/117 Calibration Information Testing the VoltAlert Function (117 only) Use this is off. 2. Connect a double banana plug to the output voltage terminals of the calibrator (Fluke 5500A or equivalent). 3. Set the calibrator output to 10 V at 60 Hz. 4. Hold the Meter so that the top
  • Fluke 116 CAL | Product Manual - Page 16
    silent and the red LED is off. 2. Connect wire (Pomona 4911A-12-2 or equivalent) to the High output voltage terminals of the calibrator (Fluke 5500A or equivalent). 3. Set the calibrator output to 10 V at 60 Hz. 4. Hold the Meter so that the wire contacts and runs across the top center of the unit
  • Fluke 116 CAL | Product Manual - Page 17
    any of the performance tests, perform the Calibration Adjustment Procedure. 110/114/115/116/117 Use these steps to view the Meter calibration counter: 1. While pressing K, turn the . 7. Press g to store the new password. MyFlukeStore 14 Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 116 CAL | Product Manual - Page 18
    calibration password is forgotten, the default password (1234) can be manually from OFF to any ON position. 4. Short across the S7 CAL keypad on the back of the PCA. See Figure 6. The Meter Calibration Password Reset Short S7 to reset to Default password XBT2 XBT1 MyFlukeStore Shop for Fluke
  • Fluke 116 CAL | Product Manual - Page 19
    the adjustment procedure, the calibration constants do not change. To adjust the Meter calibration: 1. 110/114/115/116/117: While pressing K, to indicate that the calibration adjustment is complete. 8. Press g to return to meter mode. MyFlukeStore 16 Shop for Fluke products online at: www. .com
  • Fluke 116 CAL | Product Manual - Page 20
     Table 5. Calibration Adjustment Steps (110/114/115/116/117) 110/114 Display Reading 115[1] 116[1] 117[1, 2] Input Terminals Calibrator Source Value NA C001 60 Hz 600 V, 60 Hz 6 V, 0 Hz 60 V, 0 Hz 600 V, 0Hz MyFlukeStore Shop for Fluke products online at: www. 17 .com 1.888.610.7664
  • Fluke 116 CAL | Product Manual - Page 21
    116/117 Calibration Information Table 5. Calibration Adjustment Steps (110/114/115/116/117) (cont.) Rotary Switch Position 110/114 Display Reading 115[1] 116[1] 117[1, 2] Input Terminals Calibrator Source Value Set calibrator for the 115C. [2] Do not calibrate the 117 or 117C with a line
  • Fluke 116 CAL | Product Manual - Page 22
    BRACKET MASK, 115 China FLUKE-117-2006-03,BRACKET MASK, 116 FLUKE-117-2006-05,BRACKET MASK, 116 China  FLUKE-117-8005,DIFFUSER, BACKLIGHT FLUKE-117-2001,CASE TOP, 117 FLUKE-117-2001-07,CASE TOP, 110 FLUKE-117-2001-04,CASE TOP, 113  FLUKE-117-2001-01,CASE TOP, 114 FLUKE-117-2001-02,CASE TOP
  • Fluke 116 CAL | Product Manual - Page 23
    110/113/114/115/116/117 Calibration Information Table 7. Replacement Parts (cont.) Item Description Part Number FLUKE-117-2002,CASE BOTTOM  FLUKE-117-2002,CASE BOTTOM, 11X China 2525566 2631098 FLUKE-117-2003,BATTERY DOOR,TILT STAND FLUKE-117-2003,BATTERY DOOR, TILT STAND, 11X China
  • Fluke 116 CAL | Product Manual - Page 24
    Figure 7. Exploded View of Meter 6 5 12 9 10 7 23 16 8 3 1 4 2 11 15 4 PL 13 True-rms Multimeter Replacement Parts 17 2 PL 20 21 19 14 18 22 MyFlukeStore Shop for Fluke products online at: www. 21 .com 1.888.610.7664
  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24

September 2006 Rev. 2, 5/20
©2006-2020 Fluke Corporation. All rights reserved.
All product names are trademarks of their respective companies.
110/113/114/115/116/117
True-rms Multimeter
Calibration Information
Shop for Fluke products online at:
1.888.610.7664
www.
MyFlukeStore