Table 4. DMM Performance Tests: 113, Function, Range, Applied, Display Reading

Page 14 highlights

True-rms Multimeter Performance Tests Table 4. DMM Performance Tests: 113 Step Function Range Applied Display Reading 1. 600.0 0.0 Ω 0.0 to 0.2 (2-Wire comp) 2. e 3. Ohms 600.0 6.000 k 500 Ω 5 kΩ 495.3 to 504.7 4.954 to 5.046 4. 60.00 k 50 kΩ 49.54 to 50.46 5. k CHEK 6. Continuity NA 20 Ω Beeper On NA 250 Ω Beeper Off 7. k CHEK  Diode Test NA 1.9 V 1.859 to 1.941 8. 6.000 V 0.1 V 0.095 to 0.105 9. 6.000 V 5 V 4.897 to 5.103 10. 6.000 V -5 V -5.103 to -4.897 11. 6.000 V 5 V, 45 Hz 4.897 to 5.103 12. 6.000 V 5 V, 1 kHz 4.797 to 5.203 13. 60.00 V 50 V 48.97 to 51.03 14. k CHEK Volts [1] 60.00 V -50 V -51.03 to -48.97 15. 60.00 V 50 V, 500 Hz 48.97 to 51.03 16. 60.00 V 50 V, 1 kHz 47.97 to 52.03 17. 600.0 V [2] 600 V 587.7 to 612.3 18. 600.0 V [2] -600 V -612.3 to -587.7 19. 600.0 V [2] 600 V, 45 Hz 587.7 to 612.3 20. 600.0 V [2] 600 V, 1 kHz 575.7 to 624.3 21.  Capacitance 22. 1000 nF 9999 μF Open 900 μF 0 to 2 881 to 919 [1] Manually select the range by pressing q. [2] To keep from tripping the calibrator to standby, ramp up the voltage in 50 V increments with a five-second delay between increments. MyFlukeStore Shop for Fluke products online at: www. 11 .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24

True-rms Multimeter
Performance Tests
Table 4. DMM Performance Tests: 113
Step
Function
Range
Applied
Display Reading
1.
e
Ohms
600.0
0.0
Ω
0.0 to 0.2 (2-Wire comp)
2.
600.0
500
Ω
495.3 to 504.7
3.
6.000 k
5 k
Ω
4.954 to 5.046
4.
60.00 k
50 k
Ω
49.54 to 50.46
5.
k
CHEK
Continuity
NA
20
Ω
Beeper On
6.
NA
250
Ω
Beeper Off
7.
k
CHEK
Diode Test
NA
1.9 V
1.859 to 1.941
8.
k
CHEK
Volts
[1]
6.000 V
0.1 V
0.095 to 0.105
9.
6.000 V
5 V
4.897 to 5.103
10.
6.000 V
-5 V
-5.103 to -4.897
11.
6.000 V
5 V, 45 Hz
4.897 to 5.103
12.
6.000 V
5 V, 1 kHz
4.797 to 5.203
13.
60.00 V
50 V
48.97 to 51.03
14.
60.00 V
-50 V
-51.03 to -48.97
15.
60.00 V
50 V, 500 Hz
48.97 to 51.03
16.
60.00 V
50 V, 1 kHz
47.97 to 52.03
17.
600.0 V
[2]
600 V
587.7 to 612.3
18.
600.0 V
[2]
-600 V
-612.3 to -587.7
19.
600.0 V
[2]
600 V, 45 Hz
587.7 to 612.3
20.
600.0 V
[2]
600 V, 1 kHz
575.7 to 624.3
21.
Capacitance
1000 nF
Open
0 to 2
22.
9999
μ
F
900
μ
F
881 to 919
[1]
Manually select the range by pressing
q
.
[2]
To keep from tripping the calibrator to standby, ramp up the voltage in 50 V increments with a five-second delay between increments.
Shop for Fluke products online at:
1.888.610.7664
www.
MyFlukeStore
.com