Fluke 116 CAL Product Manual - Page 14
Table 4. DMM Performance Tests: 113, Function, Range, Applied, Display Reading
View all Fluke 116 CAL manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 14 highlights
True-rms Multimeter Performance Tests Table 4. DMM Performance Tests: 113 Step Function Range Applied Display Reading 1. 600.0 0.0 Ω 0.0 to 0.2 (2-Wire comp) 2. e 3. Ohms 600.0 6.000 k 500 Ω 5 kΩ 495.3 to 504.7 4.954 to 5.046 4. 60.00 k 50 kΩ 49.54 to 50.46 5. k CHEK 6. Continuity NA 20 Ω Beeper On NA 250 Ω Beeper Off 7. k CHEK Diode Test NA 1.9 V 1.859 to 1.941 8. 6.000 V 0.1 V 0.095 to 0.105 9. 6.000 V 5 V 4.897 to 5.103 10. 6.000 V -5 V -5.103 to -4.897 11. 6.000 V 5 V, 45 Hz 4.897 to 5.103 12. 6.000 V 5 V, 1 kHz 4.797 to 5.203 13. 60.00 V 50 V 48.97 to 51.03 14. k CHEK Volts [1] 60.00 V -50 V -51.03 to -48.97 15. 60.00 V 50 V, 500 Hz 48.97 to 51.03 16. 60.00 V 50 V, 1 kHz 47.97 to 52.03 17. 600.0 V [2] 600 V 587.7 to 612.3 18. 600.0 V [2] -600 V -612.3 to -587.7 19. 600.0 V [2] 600 V, 45 Hz 587.7 to 612.3 20. 600.0 V [2] 600 V, 1 kHz 575.7 to 624.3 21. Capacitance 22. 1000 nF 9999 μF Open 900 μF 0 to 2 881 to 919 [1] Manually select the range by pressing q. [2] To keep from tripping the calibrator to standby, ramp up the voltage in 50 V increments with a five-second delay between increments. MyFlukeStore Shop for Fluke products online at: www. 11 .com 1.888.610.7664